Precision Medical 1900 Series Setting Up

Cga 540 connection oxygen conserving regulator
Hide thumbs Also See for 1900 Series:

Advertisement

Quick Links

Troubleshooting

If problem cannot be corrected, consult your Dealer/Home Care equipment provider.
PROBLEM
PROBABLE CAUSE
A. No fl ow
1. Cylinder valve closed
2. Conserver in "OFF"
position
3. Cylinder empty
4. Dial not at setting
5. Conserver not sensing
breath
B. Leaking at Cylinder
1. Missing or defective
Connection
O-ring Seal
Precision Medical, Inc.
300 Held Drive • Northampton, PA 18067
800-272-7285 • 610-262-6090 • fax: 610-262-6080
REMEDY
1. Turn on cylinder
2. Set to prescribed
setting
3. Replace cylinder
4. Set dial so indicator
points to a setting
5. Check position of
cannula in nose
Do not breathe
through mouth
1. Replace O-ring Seal
504442 (E) 4/05 Printed in USA
Setting up your
1900 Series
(CGA 540 Connection)
Quick Start Guide only
Refer to User Manual for complete operating instructions, warnings and cautions.

Advertisement

loading
Need help?

Need help?

Do you have a question about the 1900 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Precision Medical 1900 Series

  • Page 1: Troubleshooting

    PROBABLE CAUSE REMEDY A. No fl ow 1. Cylinder valve closed 1. Turn on cylinder 2. Conserver in “OFF” 2. Set to prescribed 1900 Series position setting (CGA 540 Connection) 3. Cylinder empty 3. Replace cylinder 4. Dial not at setting 4.
  • Page 2 EasyPulse5 Oxygen Conserving Regulator (Conserver) When using a Precision Medical Carry Bag, slide the cylinder into Carry Inlet the bag with the gauge facing the mesh. Attach the cannula to the O-ring Nipple Conserver’s Outlet Connection. Knob Slowly open the cylinder valve. Turn Dial to the prescribed setting.

This manual is also suitable for:

Easypulse5