Precision Medical PM5900 User Manual

Oxygen monitor
Hide thumbs Also See for PM5900:

Advertisement

U S E R M A N U A L
O
xygen
(For latest revision, go to www.precisionmedical.com)
300 Held Drive
Northampton Pa 18067 (USA)
ISO 13485 Certified
Model: PM5900
SAVE THESE INSTRUCTIONS
Federal (USA) law restricts this device to sale
by or on the order of a physician.
M
OnitOr
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
www.precisionmedical.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PM5900 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Precision Medical PM5900

  • Page 1 U S E R M A N U A L xygen OnitOr Model: PM5900 SAVE THESE INSTRUCTIONS (For latest revision, go to www.precisionmedical.com) Federal (USA) law restricts this device to sale by or on the order of a physician. Tel: (+001) 610-262-6090...
  • Page 2: Table Of Contents

    CONTENTS RECEIVING/INSPECTION ............... 2 INTENDED USE ..................2 OPERATOR PROFILE ................2 READ ALL INSTRUCTIONS BEFORE USING ........2 SAFETY INFORMATION - WARNINGS AND CAUTIONS ....... 3 SPECIFICATIONS ..................5 COMPONENT DESCRIPTION ..............6 OXYGEN MONITOR IDENTIFICATION ..........6 OXYGEN MONITOR LCD DISPLAY IDENTIFICATION ...... 8 OXYGEN SENSOR COMPONENT IDENTIFICATION .......
  • Page 3: Receiving/Inspection

    The oxygen monitor is to be used by trained healthcare professionals under the supervision, or on the order, of a physician in a hospital (or other clinical setting). The Precision Medical, Inc. Oxygen Monitor is not intended for transport use. This device is not an oxygen supply source.
  • Page 4 Follow instructions for use General Mandatory Action Sign Calibration Calibration Failed Audio Alarm Paused Low Battery Check Locked Oxygen Sensor Pass This device may contain electrical components that are hazardous to the environment. DO NOT dispose device into standard trash. Contact your local waste management for disposal of Electronic Equipment.
  • Page 5 • Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters, and Oxygen Handling. • The Oxygen Monitor should be serviced by a qualified hospital/ dealer service technician, or by Precision Medical, Inc. • DO NOT obstruct the alarm grill on the back of the Oxygen Monitor.
  • Page 6: Specifications

    SPECIFICATIONS Base Device Specifications Dimensions (Monitor without cable and sensor attached): Length: 1.72” (4.36 cm) Width: 3.56” (9.04 cm) Height: 5.44” (13.82 cm) Cable Length: 10 ft. (3.05m) (fully extended) Weight: Device Weight: 1.11 lbs (0.50 kg) (includes: monitor, sensor, batteries and cable) Shipping Weight: 1.64 lbs (0.75 kg) Operating Conditions:...
  • Page 7: Component Description

    Patient Contact: Indirect contact via gas passing through sensor sampling site. Sensor Specifications Sensor Type Precision Medical 504877 galvanic oxygen sensor (fuel cell) Expected Sensor Life > 1,000,000 O2 % Hours Specifications are subject to change without notice. COMPONENT DESCRIPTION...
  • Page 8 ITEM NAME DESCRIPTION Power Key The Power Key turns the Oxygen Monitor ON and OFF. The Lock/Unlock Key must be pressed to unlock the Oxygen Monitor, before being powered OFF. Lock / Unlock Pressing the Lock/Unlock Key will “Unlock” the keypad, so changes in the stored settings can be made.
  • Page 9: Oxygen Monitor Lcd Display Identification

    ITEM NAME DESCRIPTION Pressing the High Alarm Key when the keypad is unlocked, the high alarm set point will flash. The high alarm setting can be adjusted with the DOWN keys. When the high alarm value is changed, pressing the key will save the setting.
  • Page 10: Oxygen Sensor Component Identification

    OXYGEN SENSOR COMPONENT IDENTIFICATION Coiled Cable Locking Ring Oxygen Sensor Diveter Fitting Tee Adaptor ITEM NAME DESCRIPTION The Coiled Cable allows the Sensor to be positioned Coiled Cable with up to 10 ft from the side of the Oxygen Monitor. There Male Plugs are Male Plugs at each end...
  • Page 11: Battery Installation

    BATTERY INSTALLATION CAUTION Use ONLY AA Alkaline Batteries. 1. Press down the center, top edge of the battery cover and slide down. . Remove old batteries, if applicable. 3. Install 4 AA Alkaline Batteries, and follow the diagram stamped in the bottom of the battery compartment.
  • Page 12: Quick Setup Guide

    • Inspect the Oxygen Sensor and Diverter Fitting for visual damage or electrolyte leakage before use. DO NOT USE if damaged. • Use ONLY an Oxygen Sensor specified by Precision Medical, Inc. • The Oxygen Sensor should not be used in the presence of flammable...
  • Page 13: Calibration

    1. Calibrate the Oxygen Monitor before each use, and when replacing the Oxygen Sensor or the batteries. 2. The Precision Medical Oxygen Monitor can only be calibrated accurately using 100% Oxygen or 20.9% Oxygen (Room Air). Using any other concentration will result in inaccurate readings.
  • Page 14: Effects Of Humidity

    The following recommendation is provided to reduce the chances of pressure causing false readings. Calibrate the Precision Medical Oxygen Monitor using 100% Oxygen or room air at the same pressure and flow as the gas to be monitored. EFFECTS OF ANESTHETIC GASES:...
  • Page 15: Alarms

    Diethyl Ether or Cyclopropane. ALARMS The Precision Medical Oxygen Monitor will store the HIGH/LOW alarm settings in memory after the Oxygen Monitor is turned “OFF”. The Oxygen Monitor is designed to prevent crossing of the HIGH/LOW alarm settings.
  • Page 16: Alarm Situation

    Continuously pressing the arrow keys will move the value in increments of 5. Press the LOW ALARM SET key to save the setting. Flashing Number indicates the Number can be changed. If no keys are pressed for 10 seconds, the setting will be saved and the Oxygen Monitor will revert to locked mode.
  • Page 17: Maintenance

    Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com. DISPOSAL INSTRUCTIONS The Oxygen Monitor may contain electrical components that are hazardous to the environment.
  • Page 18: Troubleshooting

    2. Oxygen Sensor not 2. Replace with New functioning Precision Medical Oxygen Sensor 3. Using Oxygen Sensor 3. Attach New Precision other than Precision Medical Oxygen Sensor Medical Oxygen...
  • Page 19: Alarm Conditions

    Oxygen Sensor failure • Replace O2 sensor • Recalibrate O2 Monitor REPLACEMENT PARTS Description Part # User Manual 505127 Precision Medical Oxygen Sensor with Diverter 504877 Tee Adaptor 505126 Extendible Cable 504937 AA Alkaline Battery (4 Pack) 505124-4 Rubber Feet (4)
  • Page 20: Limited Warranty

    LIMITED WARRANTY LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Oxygen Monitor, (the Product), will be free of defects in workmanship and/or material for the following period: Two (2) years from shipment. Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc.