Oregon Scientific BPW810 User Manual page 15

Talking wrist type blood pressure monitor
Hide thumbs Also See for BPW810:
Table of Contents

Advertisement

Problem
Symptom
Check This
Error
Err
A measure-
shows
message
ment error
on the
occurred
display.
Err 1, 2,
The wrist
or 3
cuff is not
shows
secure
on the
display
Err 4
The monitor
shows
detected
on the
motion while
display
measuring
Err 5
Pressure is
shows
over 37.3
on the
kPa (280
display
mmHg)
Err 6
Deflation
shows
period was
on the
too long
display
Remedy
Problem
Relax for a
Error
moment and
message
then measure
again
Refasten the
cuff and then
measure again
Movement can
affect the
measurement.
Relax for a
Settings
moment and
are
then measure
wrong
again
Relax for a
moment and
then measure
again
Movement can
affect the
measurement.
Relax for a
moment and
then measure
again
14
Symptom
Check This
Retake the
EExx
A calibration
measurement. If
error
shows
the problem
occurred
on the
persists, contact
display
the retailer or our
customer service
department for
further assistance.
Refer to the
warranty for
contact
information and
return instructions
Reset the clock
Date and
The clock
time are
was not set or
incorrect
reset after
installing new
batteries
Reset the
Measure-
The
measurement
ment unit
measurement
unit
(kPa or
unit was not
mmHg)
set or reset
are
after installing
incorrect
new batteries
Remedy

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents