CareFusion Bird Sentry User Manual page 32

15625 series blender
Table of Contents

Advertisement

Chapter 7: Appendices
Mixed gas stability*
Ambient operating conditions
Temperature range
Relative humidity range
Ambient storage conditions
Temperature range
Relative humidity range
*The delivered oxygen concentration will remain constant within ±1% of the set point value with
constant inlet pressures. The displayed value may vary more than this due to sensor accuracy, age,
of the sensor, environmental conditions and length of time since last sensor calibration.
Alarm Specifications
Pressure differential alarm activation
High low concentration alarm range
O
Sensors
2
Oxygen measurement accuracy
Linearity
Potential error due to changes in
sensor temperature
Response time at 77°F (25°C) to 90%
of final reading
Ambient operating temp. range
Ambient operating humidity range
Ambient storage temp. range
Monitored gas temp. range
Monitored gas humidity range
Caution!
The use of oxygen sensors not recommended by CareFusion Corporation for use with the Bird
Sentry blender may result in performance that does not meet the specifications listed in this
manual. CareFusion Corporation recommends that only those sensors labeled as Bird
equipment be used with the Sentry.
This device has not been tested in the presence of gases other than air or oxygen.
20
Bird Sentry™ Blender
±1% oxygen
59 to 104°F (15 to 40°C)
5-95%, non-condensing
5 to 122°F (-15°C to 50°C)
5-95%, non-condensing
When supply pressures differ by 20 PSI (1.3 BAR) or more
activation
18 to 99% oxygen
±2% oxygen
±1% oxygen when temperature and pressure are held
constant
±3% oxygen, maximum
≤20 seconds
59 to 104°F (15 to 40°C)
0 to 100% (non-condensing)
50 to 122°F (–15 to 50°C)
59 to 104°F (15 to 40°C)
0 to 100%(non-condensing)
L1383 Rev. D