Bird Sentry™ Blender
User's Guide

Advertisement

Table of Contents
loading

Summary of Contents for CareFusion Bird Sentry

  • Page 1 Bird Sentry™ Blender User’s Guide...
  • Page 2 © 1997–2010 CareFusion Corporation or one of its subsidiaries. All rights reserved. The CareFusion logo and Bird Sentry are trademarks or registered trademarks of CareFusion Corporation or one of its subsidiaries. All other trademarks are the property of their respective owners.
  • Page 3 Bird Sentry™ Blender Rev ision History Date Revision Description February 1997 Initial release. March 1997 January 1998 September 2010 Rebranded the document to the CareFusion style. L1383 Rev. D...
  • Page 4: Warranty

    CareFusion, transportation charges prepaid by Buyer; (C) the defective unit or part is received by CareFusion for adjustment no later than four (4) weeks following the last day of the warranty period; and (D) examination by CareFusion of such unit or part shall disclose, to its satisfaction, that such defects or failures have not been caused by misuse, neglect, improper installation, unauthorized repair, alteration or accident.
  • Page 5: Emc Notice

    This equipment contains electromagnetic components whose operation can be affected by intense electromagnetic fields. Do not operate the Bird Sentry blender unattended in an MRI environment or in the vicinity of high- frequency surgical diathermy equipment, defibrillators, or shortwave therapy equipment.
  • Page 6: Carefusion Helpline

    Bird Sentry™ Blender Read this entire manual before attempting to operate or service the Bird Sentry™ blender: Attempting to operate the Bird Sentry blender without fully understanding its features and functions may result in unsafe operating conditions. Read the Safety Information section. It lists warnings, cautions, and notes pertinent to the safe servicing of the Bird Sentry blender.
  • Page 7: Safety Information

    The following warnings apply any time you operate or service the Bird Sentry blender: ● To avoid explosion, do not operate the Bird Sentry blender in the presence of flammable anesthetics or in an atmosphere of explosive gases. Operating the Bird Sentry blender in flammable or explosive atmospheres may result in fire or explosion.
  • Page 8 CareFusion specified lubricants when servicing the Bird Sentry blender. ● As long as the total pressure of the gas mixture being monitored is constant, the Bird Sentry blender will accurately read oxygen concentrations. However, if the total pressure varies the reading will fluctuate proportionately as the sensor actually measures the partial pressure of oxygen in the mixture.
  • Page 9 0, concentrations. The clinician should use the display panel of the Bird Sentry blender to adjust the O concentration to the desired setting. Equipment Symbols The following symbols appear on the Bird Sentry blender. Symbol Compliance Meaning...
  • Page 10 Bird Sentry™ Blender Bird Sentry™ Blender L1383 Rev. D...
  • Page 11: Table Of Contents

    Chapter 4: Troubleshooting ....................13 Chapter 5: Cleaning the Bird Sentry Blender ..............15 Chapter 6: Servicing and Maintaining the Bird Sentry Blender ..........17 Maintenance ......................... 17 Disposable Galvanic O Sensor Cable ................17 Servicing ........................18 L1383 Rev. D...
  • Page 12 Bird Sentry™ Blender Chapter 7: Appendices ...................... 19 Glossary ........................19 Specifications ........................ 19 Setup and Installation ....................21 Theory of Operations ....................22 Flow Characteristics ...................... 25 L1383 Rev. D...
  • Page 13: Chapter 1: Introduction

    Chapter 1: Introduction The Bird Sentry blender is a compact air/oxygen gas mixing device which incorporates the use of a battery powered oxygen analyzer/monitor. The gas mixing device (blender) provides for precise mixing of medical grade air and oxygen, while the analyzer measures the selected oxygen concentrations from the blender's gas flow and samples and displays these measured concentrations on a digital display.
  • Page 14: Front Panel Functional Description

    This key is used to turn on or off. The blender will continue to function and deliver mixed gas with display turned to OFF. When the ON/OFF key is pressed once, the Bird Sentry blender will start to display the oxygen concentration and the keypad is activated. If the ON/OFF key is pressed again, the Bird Sentry blender reverts to the display power off mode.
  • Page 15 UP and DOWN arrow keys. When the low alarm value is set and the LOW SET or LOCK key is pressed again, the Bird Sentry blender exits the alarm set mode and returns to normal operation. The Bird Sentry blender will also revert to normal operation (storing the most recent alarm value) if 10 seconds elapse and no keys are pressed.
  • Page 16: Bottom View

    Outlet Port A diameter indexed (DISS) oxygen fitting that, when connected with a pneumatic metering device (such as a flow meter or ventilator), delivers flows from the Bird Sentry blender from 2–100 LPM. Medical Air Inlet Connector A diameter indexed (DISS) air fitting for connection to an air-inlet hose from the air-gas source.
  • Page 17: Disposable Galvanic O Sensor

    Bird Sentry™ Blender Chapter 1: Introduction Outlet Port See Figure 1, letter O. Sensor Port A sampling port for the oxygen sensor. It allows 3 LPM of mixed gas from the blender to flow over the sensor membrane. The port is located on the back side of the plate.
  • Page 18: What You Will Need To Operate The Blender

    Do not operate the blender unless qualified personnel are in attendance to promptly respond to alarms, inoperative conditions, or sudden malfunctions. How to Get Help If you have a question regarding the installation, setup, operation, or maintenance of the Bird Sentry blender, contact customer service at the following numbers: Technical support: Hours: 7:00 a.m.
  • Page 19: Chapter 2: Operating Procedures

    Bird Sentry™ Blender Chapter 2: Operating Procedures Before placing the Bird Sentry blender into clinical use, the performance check described in Chapter 3 should be performed. Refer to Appendix C for Setup and Installation. Monitoring To begin monitoring, press the ON/OFF button located on the front panel (if necessary). Monitoring will begin immediately.
  • Page 20: Basic Operation

    To check the oxygen concentration of a sample gas: Place the sensor probe in the sensor port on the Bird Sentry blender. If using a standard T adapter in an alternate location, make sure the sensor is mounted in the adapter with the flow diverter pointing downward.
  • Page 21: Calibration Procedures

    It is best to calibrate the Bird Sentry blender using the sensor port, and with a technical grade oxygen standard (99.996% or better) at a pressure and flow similar to your application. Calibration of the unit at lower concentrations or with room air is less desirable.
  • Page 22: Automatic Calibration To Known Concentration

    Bird Sentry™ Blender Automatic Calibration to Room Air The Bird Sentry blender can be quickly calibrated to room air (20.9%) using a quick shortcut command. This function saves time by setting the calibration value to 20.9% without scrolling the display. To use this function: Place the external probe in room air.
  • Page 23: Chapter 3: Pressure Differential Alarm/Bypass Check

    Connect the 50 ±5 PSIG (3.4 ±0.3 BAR) oxygen and air supply sources to the inlet connectors located on the bottom of the Bird Sentry blender. Ensure the Bird Sentry blender is properly placed in sensor port and Bird Sentry display is on.
  • Page 24 Chapter 3: Pressure Differential Alarm/Bypass Check Bird Sentry™ Blender PROCEDURE BIRD SENTRY™ RESPONSE 11. Slowly raise the pressure of air regulator to No bubbles should be observed. Audible 50 PSIG and back to 0 PSIG while observing alarm sounds. the oxygen hose end in beaker.
  • Page 25: Chapter 4: Troubleshooting

    CareFusion Corporation for repair. 3. Bird Sentry blender operation out of calibration. Contact CareFusion distributor or CareFusion Corporation for repair. Bird Sentry blender in bypass mode, no audible Reed alarm cap damaged or defective. pressure differential alarm. Contact CareFusion distributor or CareFusion Corporation for repair.
  • Page 26 Correct temperature and recalibrate. Sensor exhibits greater than 1% oxygen 1. Sensor defective. Replace. interference in the presence of gases or vapors 2. The Bird Sentry blender requires internal listed in the sensor specifications. adjustment. Contact CareFusion distributor or CareFusion Corporation for repair.
  • Page 27: Chapter 5: Cleaning The Bird Sentry Blender

    Explosion hazard. DO NOT gas sterilize the Bird Sentry blender. The Bird Sentry blender should be kept clean and free of dust. The exterior of the Bird Sentry blender may be cleaned with a soft cloth lightly dampened with warm soapy water or mild disinfectant. An all- purpose cleaner may be used on the exterior of the Bird Sentry blender.
  • Page 28 Chapter 5: Cleaning the Bird Sentry™ Blender Bird Sentry™ Blender Bird Sentry™ Blender L1383 Rev. D...
  • Page 29: Chapter 6: Servicing And Maintaining The Bird Sentry Blender

    There are no user-serviceable components inside this device. All service should be referred to a CareFusion trained hospital/dealer service technician or CareFusion. If you have any problems operating the Bird Sentry blender, read the instructions again or consult the troubleshooting guide in Chapter 4.
  • Page 30: Servicing

    MicroBlender will be loose and is heavy. Slide the MicroBlender out from the case and two (2) screws aside for reassembly. The MicroBlender is now serviceable as a 3800 High Flow MicroBlender. The 3800 MicroBlender must be serviced by CareFusion certified technicians. Please refer to L1009 for proper cleaning and service. L1383 Rev. D...
  • Page 31: Chapter 7: Appendices

    Bird Sentry™ Blender Chapter 7: Appendices Glossary Term Description Air/ O Mixture of compressed air and oxygen °C Degrees Celsius Compressed Gas Association DISS Diameter Indexed Safety System °F Degrees Fahrenheit Fractional Concentration of Inspired Oxygen Oxygen Liters-per-minute Part number...
  • Page 32 0 to 100%(non-condensing) Caution! The use of oxygen sensors not recommended by CareFusion Corporation for use with the Bird Sentry blender may result in performance that does not meet the specifications listed in this manual. CareFusion Corporation recommends that only those sensors labeled as Bird equipment be used with the Sentry.
  • Page 33: Setup And Installation

    Bird Sentry™ Blender Chapter 7: Appendices Note: All specifications assume the following standard environmental conditions unless specified otherwise. ● Ambient and sample gas temperatures of 77°F (25°C) ● Barometric pressure of 30Hg (760mm Hg) ● Ambient relative humidity of 50% ●...
  • Page 34: Theory Of Operations

    Plug the sensor cable into the cable port on the sensor. Insert the sensor into the sensor port on the right side of the Bird Sentry blender. Allow the sensor a period of adjustment as listed in the specifications section on page 19 and stated in Chapter 3, Performance Check.
  • Page 35 Gas Outlet There are two (2) gas outlets on the Bird Sentry blender: one on the bottom of the unit and one on the left side (facing the unit). These outlets are capable of delivering combined metered flows of 2-100 liters per minute (LPM) of mixed gases.
  • Page 36 Bird Sentry™ Blender Electrical Block Diagram This diagram is intended to assist the user in understanding the general functioning of the Bird Sentry circuitry. The user and technician are both cautioned that there are no serviceable components on the circuit board. Any service requirements should be referred to CareFusion Corporation.
  • Page 37: Flow Characteristics

    Bird Sentry™ Blender Chapter 7: Appendices Flow Characteristics Note: Graphs are typical of a representative sample. Minor variations are to be expected. L1383 Rev. D...
  • Page 38 Chapter 7: Appendices Bird Sentry™ Blender Bird Sentry™ Blender L1383 Rev. D...
  • Page 39 Bird Sentry™ Blender Index gas outlet, 23 getting help, vi, 6 alarm bypass function, 23 pressure differential, 11 high alarm, setup procedure, 8 setup procedure, 7 specifications, 20 alarms and conditions, 8 installing the blender, 21 battery, installing, 21 low alarm, setup procedure, 7...
  • Page 40 Chapter 7: Appendices Bird Sentry™ Blender Bird Sentry™ Blender L1383 Rev. D...