Chapter 1: Introduction
What You Will Need to Operate the Blender
●
Pressurized Oxygen
oxygen at a line pressure of 40 to 60 PSIG (2.8 to 4.2 bar).
●
Pressurized Air
to 60 PSIG (2.8 to 4.2 bar).
What to Do If the Blender Fails
If a condition is detected that could possibly prevent the blender from continuing to operate safely, it
sounds an alarm. If at any time, ERX (Er1, Er4, etc.) appears on the LCD, contact a service technician,
certified by CareFusion.
Warning!
Do not operate the blender unless qualified personnel are in attendance to promptly respond to
alarms, inoperative conditions, or sudden malfunctions.
How to Get Help
If you have a question regarding the installation, setup, operation, or maintenance of the Bird Sentry blender,
contact customer service at the following numbers:
Technical support:
Hours: 7:00 a.m. to 4:30 p.m. (PST), Monday through Friday
Tel. 714.283.2228
(Follow the prompts to technical support.)
Fax 714.283.8471
support.vent.us@carefusion.com
CareFusion helpline:
Hours: 24 hours, 7 days a week
Tel. 800.231.2466 or 800.328.4139
6
. The compressed oxygen source must provide clean and dry, medical grade
. The air source must provide clean and dry, medical grade air at a line pressure of 40
Bird Sentry™ Blender
L1383 Rev. D