Barnstead 30400 Operation Manual page 30

Series 1262 thermolyne furnace
Table of Contents

Advertisement

8 S
& 4
16 S
EGMENT
X
Note
The program ramp rate is designed to
reduce the heatup rate or cooling rate
that the furnace normally exhibits.
When not using this feature, the fur-
nace will operate at its maximum
heating and cooling capability.
Note
When the program ramp has ended
or has been reset, the furnace will
continue to maintain setpoint temper-
ature. It will not cool to ambient tem-
perature unless the setpoint is set to
ambient temperature by the program
or by the operator.
30
P
EGMENT
ROGRAMMABLE
M
/OTP
ODELS W
dwL.U: Setting Dwell Units
Dwell units are time units which are used in "dwELL" seg-
ments (amount of time to remain at a specific temperature
). See "Setting the Segment Type" for an explanation on
how to set a dwell segment.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"dwL.U."
3.
Press the UP or DOWN button to toggle between
seconds, minutes and hours.
CYC.n: Setting the Number of Cycles
Set the number of times a group of segments or programs
are to be repeated by following the steps listed below.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display
reads,"CYC.n."
3.
Press the UP or DOWN button to select the num-
ber of cycles you want to run or, press the DOWN
button to select "cont." so the program will run
continuously.
Setting the Segment Type
There are five segment types. Proceed with the following
steps according to the type of segment you have selected.
rmP.r (Ramp)
To ramp linearly at a set rate to a specified temperature:
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display
reads,"tYPE."
3.
Press the UP or DOWN button until display
reads, "rmP.r."