Alarms - Barnstead 30400 Operation Manual

Series 1262 thermolyne furnace
Table of Contents

Advertisement

Note
The following alarm messages are
factory default settings and may vary
if you have changed the configuration
of your controller:
IDHi: = 50°C
2FSH = 1118°C
S
INGLE
Pid List
Pb: Proportional band (in display units).
ti: Integral time in seconds.
td: Derivative time in seconds.
ACCS List Code: Access code (Code needed to enter or
change the other configuration parameters which are not
normally accessible.) Not accessable.

Alarms

The controller will flash an alarm message in the home
display if an alarm condition is detected.
2FSH: Measured value full scale high alarm.
IdHi: Measured value deviation high alarm.
S.br: Sensor break: check that sensor is connected cor-
rectly.
L.br: Loop break: check that the heating circuits are work-
ing properly.
Ld.F: Heater Circuit fault: indication of either an open or
short solid sate relay, a blown fuse, missing supply or
open circuit heater.