Press the operator button to begin the test.
6.
The lift arm moves down into bin 1.
The lift arm picks up and releases the disc to measure the gripper length.
The lift arm moves up to the top position.
The lift arm moves back down into bin 1, picks up the disc, and releases the disc into bin 2 to measure the lift
length.
The carousel rotates back to bin 1.
-
Important! If you moved the lift arm to the top of the lift column in step 4, the lift arm moves to the carousel
and the operator panel displays [CAL CAROUSEL POS PUT DISC ON GRPR]. Place one disc on the lift arm and
continue to step 8 in the
The autoloader exits diagnostic test 3 and the system restarts.

Note: Diagnostic test 3 causes the autoloader to exit diagnostic mode. To continue the diagnostic tests, refer to
the
Access and Select Diagnostic Tests

Calibrate the Carousel

This test measures the position of the carousel, as well as the gear-train backlash of the carousel.
1.
Remove all discs from the carousel bins.
2.
Make sure that the autoloader is in diagnostic mode. Refer to the
3.
Press the operator button three times and hold on the third press until the operator panel displays [03 CALIBRATE].
Release the operator button. The lift arm moves to the top position and the light on the operator button flashes.
4.
Move the lift arm to the top of the lift column.
5.
Place one disc on the lift arm. Press the disc into the gripper while you press the disc release button.
6.
Press the operator button. The carousel rotates and the lift arm lowers
7.
to bin 1.
Manually, slowly lower the lift arm and the disc into the bin. The
8.
autoloader will not allow the disc to move below about 0.25 inches
from the top of the carousel.

Note: The autoloader measures the position of the carousel,
and the operator panel displays [CAL CAROUSEL POS O+00000
N+00000], where O+00000 is the old offset value and N+00000
is the new offset value based on the current position of the
carousel.
2001475_A
section on page 19.
Calibrate the Carousel
section on page 16.
Maintain the Producer III
Access and Select Diagnostic Tests
section on page 16.