Olympus FLUOVIEW FV1000 User Manual page 32

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Centration of Mercury Burner
4.
TIP
Hereafter, centering is possible with the method described in the User's Manual for the
Reflected Fluorescence System, that is, using the burner centering knob and mirror
focusing screw.
5.
TIP
The mercury burner does not have to be centered until the next time it is replaced.
II. PREPARATION For OPERATION
II. 3 - 2
Page
/ Centering the Mercury Burner
Turn the collector lens focusing knob
into focus.
After completing centering, remove the centering target and connect the light guide.
When starting observation, turn the collector lens focusing knob
brightness of the observation field.
on the lamp housing to bring the arc image
to maximize the