Environmental Specifications
Working temperature
range
Operating temperature
range
Storage temperature range
Reference temperature
range
Relative humidity
Housing
Protection
Safety
Type test voltage
EMC
Emission
Immunity
Power Supply
Functional range
Safety
Fuse
Power consumption
Memory capacity
Intervals
Events
Shop for Fluke products online at:
-10°C to +55°C
0°C to +35°C
-20°C to +60°C
23°C ± 2 K
10 to 90 %, no condensation
Robust, compact housing of CYCOLOY
IP50 per EN 60529
EN 61010-1 600 V CAT III, 300 V CAT IV
pollution degree 2, double insulation
5.2 kV AC, 50 Hz / 60 Hz, 5 s
IEC/EN 61326-1, EN 55022
IEC/EN 61326-1
88 to 660 V RMS AC absolute, 50 Hz / 60 Hz
EN 61010-1 600 V CAT III, 300 V CAT IV,
pollution degree 2, double insulation
Power supply fuse can be replaced only in service
facility. Supply can be connected in parallel to
measuring inputs (up to 660 V RMS AC).
5 W
8 MB Flash-EPROM
> 10000 intervals, > 70 days with 10 min intervals
> 13000
MyFlukeStore