Inspection And Troubleshooting; Safety Inspection Checklists - Invacare TheraPure 3752G User Manual

Recessed whirlpool tubs
Table of Contents

Advertisement

6 Inspection and Troubleshooting

If any part of the system is not functioning properly, DO NOT use the system. For parts and technical service information please contact ICCI Technical
Service at (800) 668-2337, Extension 2.
6.1

Safety Inspection Checklists

Every six months or as necessary have a qualified technician perform a thorough inspection and servicing. Routine maintenance will
extend the life and efficiency of your tub.
Initially
Ensure that the maximum water supply temperature does not exceed 110°F.
Ensure that the maximum water pressure does not exceed 55 PSI.
Verify the water flow temperature with a separate thermometer.
Visually inspect for cracks in the tub surface.
Ensure that the door handle locks the door closed.
Ensure that the safety pin properly engages the door.
Check that all labels are present and legible. Replace if necessary.
Before Each Use
Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 21.
Ensure that the maximum water fill temperature does not exceed 110°F (43° C).
Ensure that the maximum water pressure does not exceed 55 PSI.
Visually inspect for cracks in the tub surface.
Ensure that the door handle locks the door closed.
Ensure that the safety pin properly engages the door.
Every Six Months
Verify the water flow temperature with a separate thermometer.
Check that all labels are present and legible. Replace if necessary.
Clean the Water Temperature Control Valve to remove lime deposit.
Part No. 1167458
6 INSPECTION AND TROUBLESHOOTING
31
Invacare® TheraPure™ Recessed Whirlpool Tubs

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Therapure 3802g

Table of Contents