Adjust By Number - Fluke Ti300 User Manual

Thermal imagers
Hide thumbs Also See for Ti300:
Table of Contents

Advertisement

Ti200, Ti300, Ti400
Users Manual
If you set a value that is 0.60,  shows on the Imager display with this
caution:
Surfaces with an emissivity of 0.60 make reliable and consistent
determination of actual temperatures problematic. The lower the
emissivity, the more potential error is associated with the Imager's
temperature measurement calculations. This is also true even when
adjustments to the emissivity and reflected background adjustments
are performed properly.
Emissivity is set directly as a value, or from a list of emissivity values for some
common materials.
If the Display is set to Display All, you see the information about
current emissivity as = x.xx.

Adjust by Number

To set the emissivity value:
1.
Go to Measurement > Emissivity > Adjust Number.
2.
Push / to change the value.
A custom emissivity value is indicated when any value not in the standard
emissivity table is selected.
Select by Table
To select from a list of common materials:
1.
Go to Measurement > Emissivity > Select Table.
2.
Push / to highlight the material.
3.
Push  to select the material.
24
Caution:
Emissivity <0.6
(see Manual)
Note
Note