Each SMARTCELL has a wide range of choices of full scale values.
Note:
The full scale range as programmed below only affects the analog output
and the data logging full-scale output.
The display will always cover the entire operating range of the SMART-CELL.
Changing the range can affect the display resolution for some SMART-CELLs.

Changing the full scale range on EC-P2:

To Enter RANGE Mode

1.
Push power button to turn unit on. Wait for power up sequence to finish
(30 seconds).
2.
Push and hold the unnamed button (second from left) first, then Push
and hold the MENU button. Hold both buttons until the screen goes blank
(two seconds). Release both buttons and the unit is now in Programming
Mode.
3.
Push the MORE button.
4.
Push the DISP (display) button.
5.
Push the MORE button and the unit is now in the RANGE mode.

Adjusting the RANGE

6.
Push RANGE button.
7.
Push INC button (increase) or DEC button (decrease) until the display
matches the desired full scale range.
8.
Push SAVE button to store new full scale range.

To Exit RANGE Mode

9.
Push DONE button twice slowly to get back into normal operation mode.