Warning Function

4.
Push WARN button.
5.
Push FUNC (function) button.
6.
Push SELECT button while viewing the arrow in the lower left hand corner
of the screen to toggle through the following three possibilities:
7.
Push SAVE button when the DOWN arrow is displayed to store new
warning function.
8.
Push DONE button to return to Alarms mode.

To Exit CAUT / WARN / ALRM Mode

9.
Push DONE button two times slowly to get back into normal operation
mode.
The Caution and Alarm functions are programmed in the same manner, except for
step four above in which you would push either the CAUT or ALRM button.

4.2 Display Variables

The Honeywell EC-P2 has two adjustable display variables. These variables control
signal conditioning of the LCD display. Variables are stored on each SMART-CELL.
The AVG (Average) variable determines the number of one second samples to
average to calculate the displayed value. The lower the averaging value, the faster
the display will change and the more "wiggle" will be displayed. The BLANK (zero
blanking) variable controls the lowest reading that will be displayed. This is to
eliminate the misleading display of signals within the zero noise of the sensor. The
following procedure will change the AVG variable.

To Enter AVG / BLANK Mode

1.
Push power button to turn unit on. Wait for power up sequence to finish
(30 seconds).
2.
Push and hold the unnamed button first, then Push and hold the Menu
button. Hold both buttons until the screen goes blank (two seconds).
Release both buttons and the unit is now in Programming Mode.
3.
Push the MORE button.
4.
Push the DISP (display) button and the unit is now in the AVG / BLANK
mode.
16
UP arrow — Warn on increasing concentration (gas goes higher
than setpoint).
DOWN arrow — Warn on decreasing concentration (gas goes
lower than setpoint).
Blank — warning function turned OFF.
Honeywell EC-P2 Instruction and Installation Manual