In 1100/1200/1260/1290 systems, the LAN Interface Card should be installed in the
N O T E
detector (DAD, MWD, FLD, VWD, RID) due to its higher data handling rate. If no
1100/1200/1260 detector available, use the pump or the autosampler (in this order).
The LAN Interface Card is shipped with the Bootp initialization mode and will use the
N O T E
parameters (IP, Subnet Mask and Default Gateway addresses) from a Bootp server. If you
need another initialization mode or other settings, refer to "Initialization Mode
Selection" on page 22 for details before doing the next step.
6 Carefully slide the LAN Interface Card into the slot. Some pressure may
7 Disconnect your PC from the network and connect the PC network card
Be careful that you connect the LAN cable to the LAN Interface Card and NOT one of the
C A U T I O N
CAN connections. The CAN bus uses 12-Volt signals, and a misconnection to the CAN bus
may destroy network equipment on the other end of the cable.
Figure 9
G1369C LAN Interface User Manual
be necessary to properly seat the board. Tighten the screws.
to the instrument's LAN Interface Card using a Crossover Network cable
(point- to- point) or alternatives, see
CAN port for
"Hosted Module Support"
connect to CAN port of the HPLC module that hosts this
LAN interface card (if required). The CAN port of the
"hosted module" connect with the provided CAN cable to
a free CAN port on the LC system, see
Connect the LAN cable to the correct connector
Installing and Cabling the LAN Interface Card
page 11
and
LAN port control of the
module/system
page
52.