YOKOGAWA ASGW User Manual page 9

Analyzer server gateway
Table of Contents

Advertisement

Coil (00001-09999)
Data
Time setting request
Individual time setting
request
Mapping area
IM 11B06E01-01
Address
Description
00001
For all analyzers connected to ASGW
Time data should be set at the holding registers
40001- 40004 before request.
Value turns to "0" after writing.
Setting for ASIU is invalid.
00nnn
Time is set for selected GC1000MkII.
"nnn" is 100 + StnID
"nnn" (=101 - 354)
The setting for ASIU is invalid.
Value turns to "0" after writing.
01001-09000
Setting the address of coil and input relay of
GC1000MkII/ASIU.
ASGW
GC1000MkII/ASIU
ASGW
GC1000MkII/ASIU
Available is the following coils
GC1000MkII
00001:
00002:
00011-00014: Stream sequence command
00021-00026: Calibration/Validation command
00101-00131: Stream (continuous) command
ASIU:
00101-00116: On commend for DO 01-16
00201-00216: Off command for DO 01-16
Setting the other address is ignored.
This value turn to "0" after writing.
only
reads
the
input
only
writes
the
Run command
Stop command
3. Functions
relay
on
coil
on
3-3