Agilent Technologies 86100A Release Notes page 45

Agilent 86100a/b/c/d infiniium dca series & n1010a flexdca, firmware release notes (a.12.02 dca and below & a.02.50 flexdca and below)
Hide thumbs Also See for 86100A:
Table of Contents

Advertisement

• Release assertion on accessing algorithm for ACVrms. The MeasID for ACVrms and
its corresponding HelpID were not in the measurement help search list.
• Fixed issues with waveform memory dialog box.
• Fixed a defect where 40 GBit/s would not highlight in the timebase dialog, when
the user would choose it via the BitRateCtrl.
• Fixed a defect where failure occurred while bringing up saved waveform from file.
• Fixed a defect where saving screen image in .ps or .eps format causes a crash. A
data buffer wasn't being allocated large enough which caused the stack and
memory to be corrupted.
• The :DISK commands: CDIRectory and DIRectory now accept the long form of the
token WAVeforms. WAVeforms was misspelled in the yuc code.
• "Orphan" radio controls - single radio controls which are not grouped with any
other controls - now appear as text labels. These "display only" radio controls do
not respond to any user action.
• 83485B and 86106A modules were defaulting to incorrect bandwidth for optical
channels. Changed code to default optical channels to high bandwidth. The user
can still set to low bandwidth via remote interface only.
• BackLight will not go off. Fixed this defect by ensuring that the screen saver timer
only resets in the event of an actual change in mouse coordinates.
• WAV:DATA? returns extra carriage return. Carriage return only appears at the end
of the output stream.
• The title of the cal failed dialog specifies the type of cal that failed instead of a
generic 'Calibration failed' title.
• Exceeding # of display labels hangs the instrument. When the user tries to create
more than 32 labels from the remote, a dialog box is displayed. The user needs
to click on the continue button to continue. Since the front panel is locked by the
remote, the user cannot continue. The dialog was replaced with a message in the
display and a remote error.
• Default Setup was generating hardware missing errors when clock recovery
modules were installed.
• Module channels of non clock modules were not being enabled when a clock
module was installed upon default setup or module install.
• Measurement queries repeated in fast succession caused invalid result to be
returned.
• Resolved MEAS:RES returning bad data.
• Fixed issue where the electrical bandwidth for 86106A and 86109A modules are
incorrect. Changed Module script to reflect the actual bandwidth. Made changes
so that the correct IFGain is used for building the acquisition look through table.
• Remote command to create directories only accepts drive A: or User Files,
resolved.
• Remote disk commands don't permit the use of mapped drives, resolved.
• Autoscale resets power units from dBm to Watts when Quick measure is on,
resolved.
• Autoscale improved to accommodate more waveforms.
Factory Only Enhancements
• Improved timebase calibration routine by making the polynomial calibration
internal. This allows field timebase calibration, prior revisions can not be
supported
N1010A FlexDCA and 86100A/B/C/D