Fluke 192B - 196B/C - 199B/C
Users Manual

Electromagnetic Immunity

The Fluke 190 series, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined
by EN-61326-1, with the addition of the following tables.
Scope Mode (10 ms/div): Trace disturbance with VPS200 voltage probe shorted
No visible disturbance
Frequency range 10 kHz to 20 MHz
Frequency range 20 MHz to 100 MHz
Frequency range 100 MHz to 1 GHz
(*) With the 20 MHz Bandwidth Filter switched on: no visible disturbance.
With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div.
Disturbance less than 10% of full scale
Frequency range 20 MHz to 100 MHz
Test Tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale.
Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted
Disturbance less than 1% of full scale
Frequency range 10 kHz to 1 GHz
110
Table 1
E = 3V/m
2 mV/div to 100 V/div
200 mV/div to 100 V/div
500 mV/div to 100 V/div
Table 2
E = 3V/m
10 mV/div to 100 mV/div
Table 3
E = 3V/m
500mV to 1000V , 500Ohm to 30 MOhm ranges
*)