Recorder Modes - Fluke 190-104 Technical Data Manual

Fluke 190-series ii 4-channel scopemeters product datasheet
Hide thumbs Also See for 190-104:
Table of Contents

Advertisement

Recorder Modes

ScopeRecord
Roll Mode
Source and display
Bandwidth
Memory depth
Min/max values
Recording modes
Stop-on-trigger
Horizontal scale
Zoom
Memory
ScopeRecord sample rate and recording timespan
Time base range
Recorded timespan
Time/division in 'view all' mode
Glitch capture
Sample rate
Resolution
Trendplot™ Recording
Source and display
Memory depth
Ranges
Recorded time span
Recording mode
Measurement speed
Horizontal scale
Zoom
Memory
Cursor measurements – all recorder modes
Source
Dual vertical lines
190C Series
199C, 196C, 192C, 215C, 225C,
Dual or multiple input waveform storage mode, using deep memory
Input A, Input B, Dual
27,000 or more data points, each holding min/max. pair of information
Min/max values are measured at high sample rate ensuring capture and display of glitches
Single sweep, continuous roll,
Start-on-Trigger (through external),
Stop-on-Trigger (through external)
ScopeRecord mode can be stopped by an individual trigger event, or by an interruption of a
repetitive trigger signal, through any input channel (through External on 190C Series)
Ranges from full record overview to zoom in up to sample level, at any record length
Up to 2 dual input ScopeRecord waveforms can
be saved for later recall and analysis.
5 ms/div to 1 min/div
6 sec to 24 hr
50 ns
20 MS/s
200 μsec to 2 sec
Single or dual input electronic paperless chart
recorder. Plots, displays and stores meter and
scope measurements.
Any combination of measurements, made on any of the input channels
18,000 points record per input. Each recorded sample point contains a minimum, a maximum
and an average value, plus a date- and timestamp.
In view-all mode: 5 min/div to 48 hr/div (overview of total record)
Up to 22 days with a resolution of 1 minute
Continuous roll for the duration of the full
recordable timespan
Up to 64x zoom
Up to 2 TrendPlot recordings can be saved for
later recall and analysis.
Any waveform trace in any waveform display mode (Scope, ScopeRecord or TrendPlot)
Cursors may be used to identify Min, Max or Average value of any datapoint in a record, with
time between cursors, time from start or absolute time.
Any combination of inputs, up to 4 channels.
All channels sampled simultaneously.
20 MHz or 20 kHz, user selectable
Start-on-Trigger (through any channel)
Stop-on-Trigger (through any channel)
Time from start, time of day
Two multiple input ScopeRecord waveforms
can be saved internally for later recall and
analysis. Direct storage on external flash
memory drive through USB host port.
2 min/div
48 hr
250 ns
4 MS/s
4.8 sec
Multiple channel electronic paperless
recorder. Graphically plots, displays and
stores results of up to 4 automatic scope
Normal view: 5 s/div to 30 min/div
More than 22 days, with a resolution of
Continuous recording, starting at 5 s/div. with
5 automatic measurements per second or more
Time from start, time of day
Up to 64x zoom-out for full record overview,
up to 10x zoom-in for maximum detail.
Two multiple input TrendPlot records can be
saved internally for later recall and analysis.
Direct storage on external flash memory drive
ScopeMeter 190 Series II
190 Series II
190-204, 190-104
Single sweep, continuous roll,
5 ms/div ~ 2 min/div
6 sec ~ 48 hr
0.5 s/div. ~ 4 h/div
8 ns
125 MS/s
200 μsec ~ 4.8 sec
measurement over time.
102 seconds
automatic record compression
through USB host port.
Fluke Corporation 5