Performance Tests

AC Voltage Verification Test

1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, press F1 softkey for Voltmeter.
3 Press F1 to select AC voltage measurement.
4 Connect the calibrator to the instrument's input and ground terminal as
5 Select each function and range in the order shown in Table 7- 6. Apply
6 Compare the measurement results to the corresponding test limits
Table 7-6 AC Volts Verification Test
Calibrator Voltage
6 V
6 V
60 V
60 V
300 V
540 V
590 V
Agilent U1602B/U1604B User's and Service Guide
shown in Figure 7- 2 using a banana plug.
the values from the calibrator shown in Table 7- 6 below.
shown in the Table 7- 6.
Calibrator Frequency
Range
1kHz
6 V
30 kHz
6 V
1 kHz
60 V
30 kHz
60 V
50 Hz
600 V
1 kHz
600 V
10 kHz
600 V
Min Limit
Max Limit
5.928 V
6.072 V
5.808V
6.192 V
59.28 V
60.72 V
58.08 V
61.92 V
295.80 V
304.2 V
533.40 V
546.60 V
571.1 V
608.9 V
145