7
Performance Tests
The self- calibration error code carry 15 bits (0 to 14 as stated in the bit
number in Table 6- 1). The bit number is expressed in hexadecimal with
pass as "0" and fail as "1". See below for examples:
1 In all pass status, the instrument displays "OK" instead of hexadecimal
2 In all fail status, the instrument displays 0x7fff.
3 If multi- error occurred in TE3, TE4, OE2 and OE3, the hexadecimal
Agilent U1602B/U1604B User's and Service Guide
code (0x0000)
bit [15]
0000
0111
1111
0x
7
f
code "0x600C" will be displayed as self- calibration result.
0000
0110
0000
0x
6
0
0
bit [0]
1111
1111
f
f
(Hexadecimal code)
0000
1100
C
(Hexadecimal code)
153