ALARMS AND SAFETY FUNCTIONS
This unit has the alarm and safety functions shown in Table 4, and also a self-diagnostic function.
T bl 4
AI
a e
arms an
d
~
t
f
f
sa ery unc Ions
Kind of alarm
Situation
Indication
Buzzer
Safety operation
or safety
High
If the chamber temperature
Alarm lamp flashes.
Intermittent tone
Remote alarm is
temperature
exceeds the set temperature
after a delay of 15
activated after a
alarm
more than 3°C.
minutes.
delay of 15 minutes.
Low
If the chamber temperature
Alarm lamp flashes.
Intermittent tone
Remote alarm is
temperature
is lower 3°C than the set
after a delay of 15
activated after a
alarm
temperature.
minutes.
delay of 15 minutes.
Power failure
In the event of a power
Display shuts off.
-----
Remote alarm is
alarm
failure or disconnection of
activated.
power supply plug from the
outlet
Door alarm
When the door is open.
Door check lamp is lit.
Intermittent tone
-----
after a delay of 2
minutes.
Auto return
If a key operation is not
Chamber temperature is
-----
Setting mode is
performed for about 90
displayed.
canceled.
seconds in each setting
mode.
Key lock
When the key lock is ON
-----
-----
Key input is unable.
'L 1).
Thermal sensor If the thermal sensor goes
E01 and LO(E01 and HI)
-----
Compressor operates
abnormality
open (E01/LO)
are displayed alternately on the
continuously.
or short circuit (E01/HI).
temperature display.
If the defrost sensor goes
E02and chamber temperature
Normal running and
Open (E02)or short circuit
are displayed alternately on the
defrost will be
(E02).
temperature display.
passed.
Note:
The alarm can be canceled by pressing the alarm buzzer stop key (BUZZER), but the remote alarm cannot be
silenced.
After a power failure, the unit will resume operation with the set value that was in place before power failure
occurred.
DEFROSTING
The defrosting is activated when DEF key is pressed more than 5 seconds.
Remove and store the content to other freezer before starting the defrosting.
Place cotton bath towels at the
bottom of the freezer chamber before the start of defrosting as the defrost water will fall.
Keep the door open
in order to quicken defrosting.
If bath towels are soaked, remove from the chamber, dry them and replace to
prevent water to drip out of chamber.
Do not try to physically remove ice as that may cause the damage to
the evaporator.
After defrosting, remove wet towels, dry inside chamber with dry soft clothes before re-start
of the unit.
Pressing the DEF key for more than
5
seconds will restart the operation of the freezer.
16