Model NGA 2000 HFID
Preamp Assembly
656945
Case Pressure Purge
Switch 903690
Computer Analysis Board
659149
Sensor Board
657060
a.
Printed Circuit Boards
When replacing a circuit board, the fol-
lowing procedure is recommended:
Per Section 4-6, remove securing hard-
ware from electronics assembly and lay
on bench.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Safety Board
657499
Case Temperature Sensor
656026
Figure 4-12. Electronics Assembly – Exploded View
Instruction Manual
Remove securing hardware from printed
circuit board to be replaced, do not dis-
connect cable(s).
One at a time, remove the wiring con-
nectors and attach to replacement board.
Mount replacement board to electronics
assembly.
Maintenance and Service
760004-A
February 2002
Power Supply Board
655764
4-15