Campbell IRTS-P Instruction Manual page 9

Precision infrared temperature sensor
Table of Contents

Advertisement

2: Thermocouple Temp (DIFF) (P14)
1: 1
Reps
2: 21
2.5 mV 60 Hz Rejection Range
3: 2
DIFF Channel
4: 3
Type K (Chromel-Alumel)
5: 1
Ref Temp (Deg. C) Loc [ RefTemp ]
6: 2
Loc [ AppTargT ]
7: 1.0
Mult
8: 0.0
Offset
;Measure Sensor Body Temp
3: Thermocouple Temp (DIFF) (P14)
1: 1
Reps
2: 21
2.5 mV 60 Hz Rejection Range
3: 3
DIFF Channel
4: 3
Type K (Chromel-Alumel)
5: 1
Ref Temp (Deg. C) Loc [ RefTemp ]
6: 3
Loc [ SenBodyT ]
7: 1.0
Mult
8: 0.0
Offset
;Calculate P, H, & K Coefficients
4: Polynomial (P55)
1: 1
Reps
2: 3
X Loc [ SenBodyT ]
3: 4
F(X) Loc [ Psb
4: 49.9092
C0
5: 0.59237
C1
6: 0.00558
C2
7: 0.0
C3
8: 0.0
C4
9: 0.0
C5
5: Polynomial (P55)
1: 1
Reps
2: 3
X Loc [ SenBodyT ]
3: 5
F(X) Loc [ Hsb
4: 4.2828
C0
5: 0.4248
C1
6: -0.00077
C2
7: 0.0
C3
8: 0.0
C4
9: 0.0
C5
6: Polynomial (P55)
1: 1
Reps
2: 3
X Loc [ SenBodyT ]
3: 6
F(X) Loc [ Ksb
4: 52.0705
C0
5: -5.3816
C1
6: 0.387
C2
7: 0.0
C3
8: 0.0
C4
9: 0.0
C5
IRTS-P Precision Infrared Temperature Sensor
]
]
]