Chapter 4 Features and Functions
Scanning
Using External Triggering
Two control lines are provided in the rear panel mini-DIN connector:
external trigger in and external trigger out. These lines can be used
individually or combined to synchronize a scan list with an external
instrument (such as a DMM). The figure below shows this connection.
The 3499A/B/C can be configured to output a trigger pulse to notify the
external instrument whenever a channel is closed.
A g i l e n t 3 4 9 9 B
1
3
6
7
( E x t . T r i g . I n )
In addition to the Ext.Trig.In and Ext.Trig.Out pair provided on the
rear panel of the mainframe, the EI (external increment) and CC
(channel closed) pair on a 44474A module can also be used to
synchronize. Specify either Slot 0 or the slot in which a 44474A
module is installed to indicate which lines are to be used.
Both the built-in trigger lines and the EI/CC lines on the 44474A are
TTL compatible.
Once enabled, the selected trigger-in line is immediately ready to
accept the trigger signal from the external instrument. Enabling or
disabling a trigger source from the front-panel sets both the trigger in
and trigger out functions.