Agilent Technologies 4263B Operation Manual page 5

Lcr meter
Hide thumbs Also See for 4263B:
Table of Contents

Advertisement

Change 5
Add TAR in 1 m Impedance Measurement Accuracy Test (Page 9-43) as follows.
Measurement:
Cable length:
Measurement time:
Test signal level:
Status
Frequency [Hz]
Standard
100 pF
1 k
100 pF
1 k
100 pF
10 k
100 pF
10 k
100 pF
20 k
100 pF
20 k
100 pF
100 k
100 pF
100 k
1 nF
100
1 nF
100
1 μF
100
1 μF
100
1 μF
1 k
1 μF
1 k
1 μF
10 k
1 μF
10 k
1 μF
100 k
1 μF
100 k
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
Cable length:
Measurement time:
Test signal level:
Status
Frequency [Hz]
Standard
100
100 mΩ
1k
100 mΩ
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
Cable length:
Measurement time:
Test signal level:
Status
Frequency [Hz]
Standard
100 mΩ
R
100 kΩ
R
Note 1: Result = Measurement – Calibrated Value of Standard
Copyright 2009 Agilent Technologies
C
Capacitance
1 m
Long
1 V
PASS
Parameter
Limits
Cp
±0.19 pF
D
±0.0019
Cp
±0.48 pF
D
±0.0048
Cp
±1.10 pF
D
±0.0110
Cp
±1.37 pF
D
±0.0137
Cp
±0.002 nF
D
±0.0020
Cp
±0.0018 μF
D
±0.0018
Cp
±0.0011 μF
D
±0.0011
Cp
±0.0026 μF
D
±0.0026
Cp
±0.0207 μF
D
±0.0207
Resistance
1 m
Long
1 V
PASS
Parameter
Limits
R
±0.60 mΩ
R
±0.60 mΩ
DC Resistance
1 m
Long
1 V
PASS
1
Limits
Result
±1.05 mΩ
mΩ
±0.87 kΩ
mΩ
1
Result
Uncertainty
pF ± 0.0077 pF
± 0.000063
pF ± 0.012 pF
± 0.00013
pF ± 0.024 pF
± 0.000085
pF ± 0.010 pF
± 0.00016
nF ± 0.000091 nF
± 0.000063
μF ± 0.000075 μF
± 0.000047
μF ± 0.000069 μF
± 0.000033
μF ± 0.00014 μF
± 0.000061
μF ± 0.00016 μF
± 0.00029
1
Result
Uncertainty
mΩ
± 0.030 mΩ
mΩ
± 0.031 mΩ
Uncertainty
TAR
± 0.030
35.88
mΩ
26.19
± 0.014 kΩ
TAR
24.59
30.52
40.75
37.17
47.09
130.36
136.75
87.72
21.61
32.10
24.28
38.74
15.58
33.55
19.09
43.04
132.97
73.20
TAR
20.44
19.33

Advertisement

Table of Contents
loading

Table of Contents