HP E1411A Service Manual page 27

5 1/2 digit multimeter
Table of Contents

Advertisement

NOTE
Test Procedure
using CONF:VOLT:AC .07.
Chapter 4
Figure 4-3. AC Voltage Setup
Set the HP E1411 multimeter as follows:
2.
Reset Multimeter ..................................................*RST
Auto Zero ................................................................ON
Power Line Cycles (PLC) ...........................................1
Line Freq Reference (CAL:LFR) ..... 50Hz or 60Hz
*RST sets Auto Zero to ON and Power Line Cycles to 1.
Set the AC Standard Output to 0.07 Vac at 20 Hz sine wave.
1.
Set the HP E1411 range to 79.5 mV (87.5 mV with 10% overrange)
2.
Measure the AC input voltage with READ? and verify the
3.
results are within specified limits (at the range selected for 1 PLC).
Record the results on the Performance Test Record.
Repeat steps 1 through 3 using the following AC Standard
4.
voltage and frequency settings, and HP E1411 ranges:
Verification Tests 27

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1411b

Table of Contents