Test 4-3: AC Voltage Test
NOTE
Equipment Setup
WARNING
26 Verification Tests
90
FOR I=1 TO 5
100
PRINT "Set DC Standard to ";Volts(I);"VDC"
110
PRINT "Press Continue when ready"
120
PAUSE
130
CLEAR SCREEN
140
OUTPUT 70903;"CONF:VOLT:DC";Range(I) !Set DCV, range
150
WAIT 1
160
OUTPUT 70903;"READ?"
170
ENTER 70903;Read_meas(I)
180 NEXT I
190 FOR I=1 TO 5
200
PRINT "Voltage on";Range(I);"V range =";Read_meas(I);"VDC"
210 NEXT I
220 END
This test verifies AC voltage accuracy on the 87.5 mV and 300 V ranges
using sine wave inputs at 50% of full scale. The input frequency varies
from 20 Hz to 10 kHz. The DC component must be <10% of the AC
component.
The DC Voltage Performance test must be performed prior to the AC
Voltage test to check the A/D accuracy on all ranges. If the DC Voltage test
has not been performed, the AC voltage must be checked on all ranges.
Connect the equipment as shown in Figure 4-3.
1.
The AC Standard (Datron 4708, Option 20) can produce
dangerous voltages which are present on the terminals. Do not
touch the front (or rear) panel terminals unless you are sure no
dangerous voltage is present.
!Wait for settling
!Enter DC voltage