HP E1411A Service Manual page 25

5 1/2 digit multimeter
Table of Contents

Advertisement

NOTE
Test Procedure
with
settings
Example: DC Voltage
Test (DCV Input)
NOTE
Chapter 4
*RST sets Auto Zero to ON and Power Line Cycles to 1.
Set the DC Standard Output to 0.1 DCV.
1.
Set the HP E1411 range to 0.113 V (0.125 V with 10% overrange)
2.
CONF:VOLT:DC .1
Measure the input with READ? and verify the results are within
3.
specified limits (at the range selected for 1 PLC). Record results
on Performance Test Record.
Repeat steps 1 through 3 for the following DC Standard voltage
4.
and HP E1411 ranges:
E1411 Range
0.91 V
7.27 V
58.10 V
300.00 V
Remove power and disconnect test equipment.
5.
This example performs a DCV test for DC volts input and a power line
reference frequency of 60 Hz. Change line 80 to OUTPUT
70903;"CAL:LFR 50" for 50 Hz operation.
When connected to the HP E1411 multimeter, some DC Standards may
exhibit voltage variations at the start of a measurement. The WAIT 1
statement (line 150) provides a one second wait before measurement to
allow settling of the DC Standard output.
1
!DC Voltage Performance Verification
10
OPTION BASE 1
20
DIM Range(5), Volts(5), Read_meas(5)
30
DATA 0.113, 0.91, 7.27, 58.1, 300.0
40
READ Range(*)
50
DATA 0.1, 0.9, 7.0, 58.0, 300.0
60
READ Volts(*)
70
OUTPUT 70903;"*RST"
80
OUTPUT 70903;"CAL:LFR 60"
10% overrange
DC Std Output
1 V
8 V
64 V
None
300.0 V
!Set autozero on and PLC 1
!Set 60 Hz line frequency
0.9 V
7.0 V
58.0 V
Verification Tests 25