6.4.2 Forced stop of auto calibration

This mode is used to cancel the auto calibration forcedly.
1. In the Menu mode that is displayed, point the cursor
to "Setting of Auto Calibration" by pressing the UP or
DOWN key. Press the ENT key.
2. In the "Setting of Auto Calibration" item selection
screen that appears, point the cursor to "Setting of Auto
Calibration" by pressing the UP or DOWN key. Press the
ENT key.
3. "Stop Auto Calibration" is inverted. A message
appears, prompting you to verify that you want to
cancel or continue auto calibration. To cancel the auto
calibration, press the ENT key. If you press the ESC key,
auto calibration is not stopped.