5 Failure and Error Codes
to be calibrated:
a. Press SELECT.
b. Use the right or left arrow key to highlight the 1MAN option.
c. Press ENTER.
2. At the remote probe site, introduce certified span gas to the sensor.
3. Wait 60 seconds so that the reading can stabilize.
a. Note: wait 120 seconds if using A3 or A5.
4. Remove the span gas and introduce zero gas to the sensor.
5. Wait 60 seconds so that the reading can stabilize.
a. Note: wait 120 seconds if using A3 or A5.
6. At the Front Panel of the Channel Module being calibrated:
a. Press the up or down arrow key until the display reads "0".
b. Press ENTER.
c. Press the up or down arrow key until the display reads the value that
corresponds to the value of the certified span gas used.
d. Press ENTER.
7. If necessary, proceed to the previous or next procedure. Otherwise, press RUN
to place the instrument in the analysis mode.
4.5

Alarm Configuration Procedures

4.5.1 Defining the Setpoint for Alarm 1 and Alarm 2

1. At the Front Panel of the Channel Module to
configured:
a. Press SELECT twice to enter the
Configuration" Mode. ("A1" {Alarm
flashing.)
b. Use the up or down arrow key to highlight the appropriate alarm to be
defined ("A1" = Alarm 1 or "A2" = Alarm 2).
c. Press ENTER to select the alarm.
Teledyne Analytical Instruments
Model 3220
be
"Alarm
1} will be
32