Teledyne 5000B Operating Instructions Manual page 30

Photometric analyzer
Table of Contents

Advertisement

/Control Unit
3 Operation
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that
are not yet accepted by use of the Enter button.
Figure 3-1 shows the hierarchy of functions available to the operator via the
function buttons. The six functions of the buttons on the analyzer are:
Analyze. This is the normal operating mode. The analyzer monitors
the concentration of the sample, displays the percent or parts-per-
million of target fluid or contamination, and warns of any alarm
conditions.
System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
Setup an Auto-Cal
Assign Passwords
Log out to secure system
Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
Checking model and software version
Adjust electronic filter of the signal
Display more subfunctions
Display negative readings
Two of these are for programming/reprogramming the analyzer:
Define fluid applications and ranges (Refer to programming
section, or contact factory.)
Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
Zero. Used to set up a zero calibration.
Span. Used to set up a span calibration.
Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.
Range. Used to set up three analysis ranges that can be switched
automatically with autoranging or used as individual fixed
ranges.
Any function can be selected at any time by pressing the appropriate
button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the following proce-
dures. The VFD screen text that accompanies each operation is reproduced, at
3-2
Part I
Teledyne Analytical Instruments