Teledyne 5000B Operating Instructions Manual
Teledyne 5000B Operating Instructions Manual

Teledyne 5000B Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Quick Links

Photometric Analyzer

OPERATING INSTRUCTIONS

5000B
Model
Photometric Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
M5000B
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
11/02/2006
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
i
Teledyne Analytical Instruments

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne 5000B

  • Page 1: Operating Instructions

    HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE 11/02/2006 CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER. Teledyne Analytical Instruments...
  • Page 2 Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
  • Page 3 Photometric Analyzer Table of Contents Part I: Control Unit ........ Part I: 1-1 Part II: Analysis Unit ......Part II: 4-1 Appendix ............A-1 Teledyne Analytical Instruments...
  • Page 4 Model 5000B Teledyne Analytical Instruments...
  • Page 5 Part I: Control Unit OPERATING INSTRUCTIONS 5000B Model Photometric Analyzer Part I: Control Unit NEMA 4 Bulkhead Mount Teledyne Analytical Instruments Part I: i...
  • Page 6: Table Of Contents

    Model 5000B Photometric Analyzer Table of Contents 1 Introduction 1.1 Overview ................ 1-1 1.2 Typical Applications ............1-1 1.3 Main Features of the Analyzer ........1-5 1.4 Operator Interface ............1-6 1.4.1 Up/Down Switch ............ 1-6 1.4.2 Escape/Enter Switch ..........1-6 1.5 Control Section Interface Panel........
  • Page 7 3.8.2 The Curve Algorithm Screen ......... 3-31 3.8.2.1 Manual Mode Linearization ......3-31 3.8.2.2 Auto Mode Linearization ........ 3-31 4 Maintenance 4.1 Fuse Replacement............4-1 4.2 System Self Diagnostic Test ........... 4-2 4.3 Major Internal Components ..........4-3 Part I: iii Teledyne Analytical Instruments...
  • Page 8 Model 5000B Photometric Analyzer iv: Part I Teledyne Analytical Instruments...
  • Page 9: Introduction

    5000B Analysis Unit, is versatile microprocessor-based instrument. Part I, of this manual covers the Model 5000B General Purpose NEMA 4 Bulkhead Mount Control Unit. (The Analysis Unit is covered in Part II of this manual.) The Control Unit is for indoor/outdoor use in a nonhazardous environment only.
  • Page 10 1 Introduction Model 5000B Ethanol Isopropanol Methanol Alkanes, including: 0–500 ppm Heptane Hexane Ammonia 0–1000 ppm and up Aromatics, including: 0–500 ppm Benzene Cumene Toluene Xylene Chlorinated Hydrocarbons including: 0–200 ppm Carbon Tetrachloride Ethyl Chloride Ethylene Dichloride Methyl Chloride Perchloroethylene...
  • Page 11 0–15% Sulfur Dioxide 0–1000 ppm Vinyl Acetate 0–2% NOTE:Range may be higher or lower per application. OTHER NIR ABSORBERS (that can be measured) (consult factory) Acetic Acid Alcohols Amines Aromatics Butadiene Carbonyls Chloroprene Esters Part I: 1-3 Teledyne Analytical Instruments...
  • Page 12 1 Introduction Model 5000B Hydroxyl Value Hydrocarbons Hydrogen Chloride Hydrogen Fluoride Hydroxyl Value Ketones Olefins Oximes Epoxides Methylene Ketones 1-4: Part I Teledyne Analytical Instruments...
  • Page 13: Main Features Of The Analyzer

    Photometric Analyzer Part I: Control Unit Main Features of the Analyzer The Model 5000B Photometric Analyzer is sophisticated yet simple to use. The main features of the analyzer include: • A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator.
  • Page 14: Operator Interface

    1 Introduction Model 5000B Operator Interface All controls and displays on the standard 5000B are accessible from outside the housing. The instrument has two simple operator controls. The operator has constant feedback from the instrument through an alphanumeric display, and a digital LED meter. The displays and controls are described briefly here and in greater detail in chapter 3.
  • Page 15 Digital Meter Display: The meter display is a Light Emitting Diode LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It is accurate across all analysis ranges. The 5000B models produce continuous readout from 0-10,000 ppm and then switch to continuous percent readout from 1-100 %.
  • Page 16: Control Section Interface Panel

    Not implemented at this printing. Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 5000B is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. 1-8: Part I...
  • Page 17 Photometric Analyzer Part I: Control Unit Figure 1-2: Model 5000B Interface Panel Part I: 1-9 Teledyne Analytical Instruments...
  • Page 18 1 Introduction Model 5000B 1-10: Part I Teledyne Analytical Instruments...
  • Page 19: Installation

    Photometric Analyzer Part I: Control Unit Installation Installation of Model 5000B Analyzers includes: 1. Unpacking, mounting, and interconnecting the Control/Analysis Section 2. Making gas connections to the system 3. Making electrical connections to the system 4. Testing the system. This chapter covers installation of the Control Section. (Installation of the Analysis Section is covered in Part II of this manual.)
  • Page 20: Primary Input Power

    1 Introduction Model 5000B Figure 2-3: Interface Panel of the Model 6600 Control Section For safe connections, ensure that no uninsulated wire extends outside of the terminal blocks. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should come in contact with fingers, tools or clothing during normal operation.
  • Page 21 0.25 V = Range 1, 0.5 V = Range 2, 0.75 V = Range 3. 4–20 mA dc % Range: Current increases linearly with increasing sample concentration, from 4 mA at 0% to 20 mA at full scale 100%. (Full scale = 100% of programmed range.) Part I: 2-3 Teledyne Analytical Instruments...
  • Page 22 1 Introduction Model 5000B 4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA = Range 3. Figure 2-5: Analog Output Connections Examples: The analog output signal has a voltage which depends on the sample concentration AND the currently activated analysis range.
  • Page 23 Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 ampers at 250 V AC into a resistive load (Figure 2-6). Part I: 2-5 Teledyne Analytical Instruments...
  • Page 24 1 Introduction Model 5000B Normally closed Normally open Normally open Moving contact Moving contact Figure 2-6: Types of Relay Contacts The connectors are: Threshold Alarm 1: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below thresh old).
  • Page 25 (See Remote Calibration Protocol below.) Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 5000B Analyzer, the customer's controller must monitor the Cal Relay Contact. When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.
  • Page 26 1 Introduction Model 5000B When CRC opens again, zero and span are done, and the sample is being analyzed. Note: The Remote Bench terminal strip (section 3.6 Part III) provides signals to ensure that the zero and span gas valves will be controlled synchronously.
  • Page 27: Testing The System

    Toggling input. Stops/Starts any status message output from the RS-232, Until st<enter> is sent again. The RS-232 protocol allows some flexibility in its implementation. Table 2- 5 lists certain RS-232 values that are required by the 5000B. Table 2-5: Required RS-232 Options Parameter...
  • Page 28: Introduction

    1 Introduction Model 5000B • Check that sample pressure typically between 0 and 30 psig, according to the requirements of your process. Power up the system, and test it by performing the following operation: 1. Repeat the Self-Diagnostic Test. 2. Zero the instrument.
  • Page 29: Using The Data Entry And Function Buttons

    Operation /Control Unit 3 Operation Introduction Although the Model 5000B is usually programmed to your application at the factory, it can be further configured at the operator level, or even, cautiously, reprogrammed. Depending on the specifics of the application, this might include all or a subset of the following procedures: •...
  • Page 30 /Control Unit 3 Operation Model 5000B The Enter button is used to accept any new entries on the VFD screen. The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button.
  • Page 31: Operation /Control Unit

    Application/Range Enter Select range Enter Span Enter gas value OUTPUT: Set current Enter output 4 or 20 MA MORE Display Enter MODEL Model/Version Show Negative SHOW.NEG Reading Figure 3-1: Hierarchy of System Functions and Subfunctions Teledyne Analytical Instruments Part I...
  • Page 32: The System Function

    /Control Unit 3 Operation Model 5000B the appropriate point in the procedure, in a Monospaced type style. Push-button names are printed in Oblique type. The System Function The subfuctions of the System function are described below. Specific procedures for their use follow the descriptions: •...
  • Page 33 If instrument is turned off, the next time the instrument is powered, the instrument will automatically perform a calibration cycle after 3 minutes of entering the sample mode if AUTOCAL functions were on prior to shut down. Teledyne Analytical Instruments Part I...
  • Page 34: Password Protection

    /Control Unit 3 Operation Model 5000B 3.3.2 Password Protection Before a unique password is assigned, the system assigns TAI by default. This password will be displayed automatically. The operator just presses the Enter key to be allowed total access to the instrument’s features.
  • Page 35: Installing Or Changing The Password

    DÑ arrow keys to change the letters to the new password. The full set of 94 characters available for pass- word use are shown in the table below. Characters Available for Password Definition: ¥ ® " & < > Teledyne Analytical Instruments Part I...
  • Page 36: Logging Out

    /Control Unit 3 Operation Model 5000B When you have finished typing the new password, press Enter. A verifica- tion screen appears. The screen will prompt you to retype your password for verification. Enter PWD To Verify: A A A Use the arrow keys to retype your password and press Enter when finished.
  • Page 37: System Self-Diagnostic Test

    Refer to Chapter 5 Maintenance and Troubleshooting for number-code information. The results screen alternates for a time with: Press Any Key To Continue... Then the analyzer returns to the initial System screen. Teledyne Analytical Instruments Part I...
  • Page 38: The Model Screen

    /Control Unit 3 Operation Model 5000B 3.3.5 The Model Screen Move the < > arrow key to MORE and press Enter. With MODEL blinking, press Enter. The screen displays the manufacturer, model, and software version information. 3.3.6 Checking Linearity with ALGORITHM From the System Function screen, select ALGORITHM, and press Enter.
  • Page 39: Digital Flter Setup

    90% response time on the different settings to a step input is shown below. This response time does not include the contributions of the bench sampling system and the preamplifier near the detector. Setting 90% Response time (seconds) Teledyne Analytical Instruments Part I 3-11...
  • Page 40: Filter Or Solenoid Setup

    /Control Unit 3 Operation Model 5000B 14.0 25.0 46.0 90.0 At a setting of “zero”, the response time is purely set by the electronics to 4.5 seconds. The numbers above can and will change depending on application and they merely serve to illustrate the effect of the digital filter.
  • Page 41: Hold/Track Setup

    The Sample timer lets the operator adjust the time the instrument purges sample fluid after finishing a calibration before it lets the analog outputs and display track the change in concentration. Teledyne Analytical Instruments Part I 3-13...
  • Page 42: Analog 4 To 20 Ma Output Calibration

    /Control Unit 3 Operation Model 5000B This function and the TRACK/HOLD feature will prevent false alarms while performing remote or autoscheduled calibrations. These functions are not applicable if the calibration is initiated through the front panel. To enter the Calibration/Hold Timer function, you must:...
  • Page 43: Model

    By using a digital Volt meter on the 0-1 Volt output. 3.3.12 Model This selection in the System menu flashes for a few seconds the model number and the software version installed in this instrument. Teledyne Analytical Instruments Part I 3-15...
  • Page 44: Show Negative

    /Control Unit 3 Operation Model 5000B 3.3.13 Show Negative The analyzer defaults to not to show negative readings on the analyze mode only. This affects the analog outputs too by pressing the UP or DOWN key, the analyzer can be set to display negative readings, on the SHOW_NEG field of the system menu.
  • Page 45: Manual Mode Zeroing

    Press Enter to begin the zero calibration. After a few seconds the first of three zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a prede- termined rate. Teledyne Analytical Instruments Part I 3-17...
  • Page 46: Cell Failure

    /Control Unit 3 Operation Model 5000B ####.## % SO2 Zero adj:2048 C—Zero The analyzer goes through C–Zero, F–Zero, and S–Zero. During C– Zero and F–Zero, use the DÑ keys to adjust displayed Zero adj: value as close as possible to zero. Then, press Enter.
  • Page 47: Span Cal

    3.4.2.2 Manual Mode Spanning Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Select span mode: MANUAL Teledyne Analytical Instruments Part I 3-19...
  • Page 48: Offset Function

    /Control Unit 3 Operation Model 5000B Use the DÑ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen. Span Val: 2Ø.ØØ % <ENT> To begin span Use the <...
  • Page 49 3. Flow a zero level of the process background fluid mix through the 5000BF sample cell on the Analyse mode. Wait for the reading to become stable. Write the reading down. Change the sign of the reading: This is the Teledyne Analytical Instruments Part I 3-21...
  • Page 50: The Alarms Function

    /Control Unit 3 Operation Model 5000B offset to be entered. 4. Do a manual run to check. Reintroduce the zero calibration fluid. Start a zero on the analyser but this time enter the offset value. 5. At the end of the zero function, check that the instrument reads the entered offset.
  • Page 51 AL1: 1ØØØ ppm HI Dft:N Fs:N Ltch:N Five parameters can be changed on this screen: • Value of the alarm setpoint, AL1: #### • Out-of-range direction, HI or LO • Defeated? Dft:Y/N (Yes/No) • Failsafe? Fs:Y/N (Yes/No) Teledyne Analytical Instruments Part I 3-23...
  • Page 52: The Range Select Function

    /Control Unit 3 Operation Model 5000B • Latching? Ltch:Y/N (Yes/No). • To define the setpoint, use the < > arrow keys to move the blinking over to AL1: ####. Then use the DÑ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing.
  • Page 53: Auto Screen

    They continue to read beyond the full- scale setting until amplifier saturation is reached. Below amplifier saturation, the overrange readings are accurate UNLESS the application uses lineariza- tion over the selected range. Teledyne Analytical Instruments Part I 3-25...
  • Page 54: Precautions

    /Control Unit 3 Operation Model 5000B The concentration ranges can be redefined using the Range function Manual screen, and the application fluides can be redefined using the System function, if they are not already defined as necessary. CAUTION: Redefining applications or ranges might require relineariza- tion and/or recalibration.
  • Page 55 Photometric Analyzer Operation /Control Unit 3 Figure 3-2 illustrates these schemes graphically. 0.01 Figure 3-2: Examples of Autoranging Schemes Teledyne Analytical Instruments Part I 3-27...
  • Page 56: The Analyze Function

    /Control Unit 3 Operation Model 5000B The Analyze Function Normally, all of the functions automatically switch back to the Analyze function when they have completed their assigned operations. Pressing the Escape button in many cases also switches the analyzer back to the Analyze function.
  • Page 57: The Set Range Screen

    1 V dc (20 mA at the current output). However, the digital readout and the RS-232 output of the concentra- tion are unaffected by the fixed range. They continue to read beyond the full- Teledyne Analytical Instruments Part I 3-29...
  • Page 58 /Control Unit 3 Operation Model 5000B scale setting until amplifier saturation is reached. Below amplifier saturation, the overrange readings are accurate UNLESS the application uses linearization over the selected range. To program the ranges, you must first perform the four steps indicated at the beginning of section 3.8 Programming.
  • Page 59: Appendix

    After each point is entered, the data-point number increments to the next point. Moving from the lowest to the highest concentration, use the DÑ keys to set the proper values at each point. Dpt INPUT OUTPUT 0 Ø.ØØ Ø.ØØ Teledyne Analytical Instruments Part I 3-31...
  • Page 60: Auto Mode Linearization

    /Control Unit 3 Operation Model 5000B Repeat the above procedure for each of the data points you are setting (up to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The linearizer will automatically adjust for the number of points entered.
  • Page 61: Maintenance

    MANUAL. Fuse Replacement The 5000B requires two 5 x 20 mm, 6.3 A, F type (Fast Blow) fuses. The fuses are located inside the main housing on the Electrical Connector Panel, as shown in Figure 4-3. To replace a fuse: 1.
  • Page 62: System Self Diagnostic Test

    3 Maintenance Model 5000B Photometric Analyzer Figure 4-1: Removing Fuse Block Cap and Fuse from Housing 2. Replace fuse by reversing process in step 1. System Self Diagnostic Test NOTE: Always run self diagnostic with the intended zero fluid. 1. Press the System button to enter the system mode.
  • Page 63: Major Internal Components

    Major Internal Components The major components in the Control Unit are shown in Figure 4-3. Figure 4-3: Control Unit Major Internal Components Part I: 4-3 Teledyne Analytical Instruments...
  • Page 64 WARNING: HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. The 5000B Control Units contain the following major components: • Power Supply • Motherboard (with Microprocessor, RS-232 chip, and Preamplifier PCB) •...
  • Page 65 Photometric Analyzer Part II: Analysis Unit NEC or ATEX Type 5000B - GP, Rack, Panel (Integral or Remote) 5000 - GP, Bulkhead (Z-Purged in Div II or Zone 2 areas) I, II, B, C, D (Integral) 5000B-(X-Purged, 1,1, B, C, D) (Integral)
  • Page 66 Model 5000B Photometric Analyzer Table of Contents 4 Operational Theory 4.0 Operations ..............4-1 4.1 Control Function ............4-2 4.1.1 Analysis Section ............ 4-2 4.1.2 Explosion-Proof Version: Control Module ....4-2 4.2 Start Up................4-3 4.2.1 Preliminary Inspection ........... 4-3 4.2.2 Pre-Start Up Electrical Checkout ......
  • Page 67 Part II: Analysis Unit Appendix A-1 Specifications ..............A-1 A-2 Recommended 2-Year Spare Parts List ......A-4 A-3 Drawing List ..............A-5 Part II: iii Teledyne Analytical Instruments...
  • Page 68 Model 5000B Photometric Analyzer iv: Part II Teledyne Analytical Instruments...
  • Page 69: Operations

    In any case, it is important that you calibrate the analyzer when it is first installed. Zero checks should be made routinely once or twice a week and span calibration should be performed weekly, at first, and monthly after reproducibility is assured. 4–1 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 70: Control Function

    These switches have different movements to achieve the same functionality as the 4 button model. (Consult factory or assure the manual has the correct version of the front panel control switches and functions. 4–2 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 71: Start Up

    Accessory test equipment is not necessary for start-up of the 5000B.. However, if the analyzer malfunctions at start-up, an oscilloscope and a multimeter will be required for troubleshooting .
  • Page 72: Pre-Start Up Electrical Checkout

    2. Before opening any of the enclosure covers with the power on, make certain that the area has been classified as safe to do so. 4–4 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 73 The electrical circuits take from 30 to 40 minutes to stabilize. NOTE: It will take at least four hours for the analyzer to completely stabilize with respect to temperature. 4–5 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 74: Calibration

    If the analyzer has been in use for some time, it is possible that the sample cell windows need cleaning or that a filter has deteriorated. Another cause of peak imbalance might be that the sample 4–6 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 75 9. Adjust the control module SPAN control to obtain the correct meter reading with reference to the concentration of span fluid. For example, if the sample contains 100% of whatever the 4–7 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 76 The absorber is actuated by a switch(s) located on the readout panel of the control unit. The option functions by pre- senting a calibrated and reproducible absorbance or fluid into the light path of the analyzer. 4–8 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 77 The on-line analyzer is only as accurate 4–9 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 78 Zero calibration using secondary standard zero and span flag(s) AFTER the analyzer has been calibrated on the process fluid, a simulation of this process on-line calibration can be made using internal synthetic 4–10 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 79 (This is usually done with N2 assuming the process fluid has low vapor pressure and can be purged dry in an acceptable short time 4–11 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 80 60-90% of the full scale range. NOTE: The zero or span flag readings should always repeat if the analyzer remains in calibration. Any drift in the zero or span can 4–12 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 81 SUMMARY The zero and/or span calibration of 5000 series photometric Teledyne analyzers can be quickly verified manually by using Zero and/or calibration zero offsets and/or span flags. Refer to typical Teledyne Analytical 1475 sample system below, 3-way valves 1,2,+ 3 below are used to check internal calibration using zero and/or span filters with N2.
  • Page 82 80 ppm simulated sensitivity change of water. Note: the span flag checks the sensitivity change of the 5000B and not the zero drift of the analyzer. The mole sieve dried TDI checks the zero of the instrument provided it can be reproduce the zero by consistently removing all the water to the same ppm level, (somewhere in the 2-10 ppm water range).
  • Page 83: Teledyne Analytical Instruments

    Span in Zero in ITEM PART NO. DESCRIPTION THIS DRAWING IS THE PROPERTY OF TELEDYNE ANALYTICAL INSTRUMENTS AND CONTAINS CONFIDENTIAL TOLERANCE UNLESS INFORMATION. IT IS NOT TO BE COPIED WITHOUT WRITTEN PERMISSION. OTHERWISE SPECIFIED TELEDYNE Analytical Instru- SCALE FRAC.
  • Page 84 10 Teledyne is not responsible for applying a general purpose instrument in a hazardous area or where a flammable or toxic gas or liquid is brought to an analyzer above its lower explosive limit and the area has been classified as general purpose and safe.
  • Page 85: Operational Theory Continued

    At this point the reference signal is fed back to the automatic gain control loop to maintain the desired system gain. In addition, both the mea- suring and reference levels are fed to selector switches in order to enable 4–17 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 86: Source Module

    CaF2 for longer wavelengths. Also the source lamp may be a differenct type, to emit more radiation at the longer measuring wavelengths. (Consult manual addendum descriptions). 4–18 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 87: Sample Cell

    1. A preheater (heat exchanger coil) is used on the incoming sample stream to raise or lower it to the desired temperature controlled level. 4–19 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 88 3. The sample is temperature compensated for by software while inputing the temperture value at the sample cell. Figure 4-2 Analyzer System - Block Diagram 4–20 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 89 T hermistor T emp. Heater Control S ample T hermistor Module Heater T emp. Control Detector T hermistor Module Heater T emp. Control Figure 4-3 Power Module - Block Diagram 4–21 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 90: Power

    1800 RPM. The filter wheel performs two func- tions: (1) switching filters, and (2) chopping the optical signal to give pulses which can be amplified for high quality processing. 4–22 Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies...
  • Page 91: Routine Maintenance

    1amp off condition occurs System Visual Check and Response Procedure Verify that the signal failure alarm is not in failure condition. Verify that the zero and span control setting have not been disturbed. Part II: 5-1 Teledyne Analytical Instruments...
  • Page 92: Suggested Preventive Maintenance Schedule

    5 Maintenance Model 5000B Photometric Analyzer Verify that the chart recorder contains a normal display. Verify that the recorder has a sufficient supply of chart paper and ink. Routine Maintenance Keep the sample lines and components, including the measuring cell within the analyzer sample module, free of deposits and leaks.
  • Page 93: Service Procedures And Adjustments

    Fill the sample cell with air or a stable fluid, such that the photo energy that strikes the detector is constant. A stable fluid is distilled or tap water. This step may be omitted when the system is stable in its present state. Part II: 5-3 Teledyne Analytical Instruments...
  • Page 94: Oscilloscope Display Of The I To E Converter Output

    5 Maintenance Model 5000B Photometric Analyzer If you open the detector module, keep stray light out by covering the opening with a dense black cloth. If you do not take this precaution, the result is a misinterpretation of the scope patterns. On general-purpose systems, the scope test points are in the bottom of the detector module and are accessible without opening the module.
  • Page 95 If the measuring peak is equal to or within 1 volt of thereference peak, the system is optically balanced and ready for calibration. c. If the peak is still too short, repeat the procedure, but thistime put a screen behind the measuring filter to shorten its peak. Part II: 5-5 Teledyne Analytical Instruments...
  • Page 96: Setup Of The Logarithmic Amplifier

    5 Maintenance Model 5000B Photometric Analyzer 7. After the peaks are balanced, adjust the gain control until the tallest of the two peaks is 8 to 9 volts. The peaks should still be within1 volt of each other. 8. It is always good practice to operate the analyzer with as low a gain as possible.
  • Page 97: Integrated Reference And Measuring Signals

    High DC voltage for the photodetector, and signals to control calibration solenoids and filters. To gain access to this terminals, the silkscreen cover must be removed. These terminals are wired in the factory. Part II: 5-7 Teledyne Analytical Instruments...
  • Page 98 5 Maintenance Model 5000B Photometric Analyzer WARNING: DANGEROUS HIGH VOLTAGES ARE PRESENT AT THESE TERMINALS. TRAINED PERSONNEL MUST REMOVE THE SILKSCREEN COVER ONLY. EXER- CISE EXTREME CAUTION. The first strip terminal has three contacts labeled N, G and H. The labels stand for Neutral, Ground, and Hot.
  • Page 99 COM: Common reference to the +/- 15 VDC and the -230 VDC power supplies. SPAN FLTR: Span filter signal, AC voltage. SPAN SOL: Span solenoid signal, AC voltage. ZERO FLTR: Zero filter signal, AC voltage. ZERO SOL: Zero solenoid signal, AC voltage. Part II: 5-9 Teledyne Analytical Instruments...
  • Page 100 5 Maintenance Model 5000B Photometric Analyzer Teledyne Analytical Instruments 5-10: Part II...
  • Page 101 Photometric Analyzer, 5000B Appendix Appendix A-1 Specifications 5000B Digital Control Module: Ranges: Four Programmable Ranges, field selectable within limits (application dependent) and Auto Ranging Display: 2 line by 20 alphanumeric VFD accompanied by 5 digit LED display Signal Output: Two 0-1V DC (concentration and range ID)
  • Page 102 Analog Output Signal: 0-1 V (standard), 4-20mA isolated Area Classification Non-hazardous:Model 5000B Hazardous / Explosion-proof: Model 5020 5000B, 5020B Purge options:X, Y, Z purge, Cenelec IIB AND H2 (PENDING) FOR ZONE I OR CLASS I, DIV I, B, C, D AREAS Teledyne Analytical Instruments...
  • Page 103 Photometric Analyzer, 5000B Appendix Enclosure Purging: Purge air startup*:40 – 80 PSI (3 – 6 BAR), 0.5 CFM (15 L / min) Purge optical path*:15 – 30 PSI (1 – 2 BAR), 10 cc/min, purity of Nitrogen, - 110 °F (-80 °C) dewpoint...
  • Page 104 Appendix Models 5000B A-2 Recommended 2-Year Spare Parts List QtyP/NDescription C-75825A Motherboard, Control Unit C-67990 Amplifier, Control Unit D-67990 5000B Interface PCB A-9306 Differential Power Supply C-40265A Measuring PCB L323 UV LAMP, 5000B B77609 Parabolic Reflector Assy. B74423 Detector assy (C78194 assy)
  • Page 105 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 961-2538 TWX (910) 584-1887 TDYANYL COID Web: www.teledyne-ai.com or your local representative. A-3 Drawing List, 5000B SEE MANUAL ADDENDUM OF Q FOR COMPLETE DRAWING LIST. Teledyne Analytical Instruments...
  • Page 106 Appendix Models 5000B Teledyne Analytical Instruments...

Table of Contents