DH Instruments HPMS A70M Operation And Maintenance Manual page 101

Reference pressure monitor with high pressure mounting system
Table of Contents

Advertisement

READYCKn
Purpose
Read or set the Ready check flag.
Command
"READYCKn 1"
Query
"READYCKn?"
Classic
"READYCKn=1"
"READYCKn"
Optional Suffix
"n"
The active Q-RPT is assumed if no suffix is given.
'1'
'2'
Remarks
The internal ready check flag is cleared whenever the specified or active Q-RPT reaches a
Not Ready (NR) condition. The "READYCK" query returns the status of the flag. The flag is
set by sending the "READYCK 1" program message while the Q-RPT is in a Ready condition.
The "READYCK" program message query can then be used at a later time to determine
whether a Not Ready condition has occurred since the ready check flag was set.
Example
Cmd sent:
(enhanced)
Query reply:
Example
Cmd sent:
(enhanced)
Query reply:
Example
Cmd sent:
(enhanced)
Query reply:
Example
Sent:
(classic)
Query reply:
Example
Sent:
(classic)
Query reply:
Errors
ERR# 6:
See Also
"SR"
3.2.1
REMOTE
Purpose
Lock out the front panel keypads during remote operation.
Command
"REMOTE"
Classic
"REMOTE"
Remarks
The RPM4 goes into remote mode whenever communications take place. The user can
return to local operation by pressing the [ESC] key. The REMOTE program message locks
out the front panel completely. The only way to unlock the front panel after the "REMOTE"
command is using the "LOCAL" program message, the IEEE-488 "GTL" command, or by
cycling the RPM4 power.
Example
Cmd sent:
(enhanced)
Reply:
Example
Cmd sent:
(enhanced)
Reply:
Example
Sent:
(classic)
Reply:
See Also
"LOCAL"
RESn
Purpose
To read or set the pressure display resolution for the active range.
Command
"RESn res"
Query
"RESn"
Classic
"RESn= res"
"RESn"
Default
"RESn=0.001"
Optional Suffix
"n"
The active Q-RPT is assumed if no suffix is given.
'1'
'2'
Arguments
res:
Remarks
The pressure display resolution is defined as % span of the active range. The setting is
separate for each range, and changes as the range is changed.
Example
Cmd sent:
(enhanced)
Query reply:
Example
Cmd sent:
(enhanced)
Query reply:
Example
Sent:
(classic)
Reply:
Errors
ERR# 6
See Also
3.4.2
Specify the Hi Q-RPT
Specify the Lo Q-RPT
"READYCK1 1"
"1" (no reply if GPIB-488)
"READYCK1?"
"1"
"READYCK?"
"1" (if Q-RPT condition has stayed Ready)
"0" (if Q-RPT condition has NOT stayed Ready)
"READYCK=1"
"READYCK=1"
"READYCK"
"READYCK=1" (if Q-RPT condition has stayed Ready)
"READYCK=0" (if Q-RPT condition has NOT stayed Ready)
Argument is not a '0' or a '1'.
"REMOTE"
"REMOTE" (no reply if IEEE-488)
"REMOTE?"
"REMOTE"
"REMOTE"
"REMOTE"
Specify the Hi Q-RPT
Specify the Lo Q-RPT
The pressure display resolution in % span of the specified Q-RPT range (0.0001 to 1 %
FS).
"RES .01"
"0.01" (No reply if GPIB-488)
"RES? .01"
"0.01"
"RES=.01"
"0.01"
The argument is invalid.
Page 91
4. REMOTE OPERATION
© 2007 DH Instruments, a Fluke Company