DH Instruments HPMS A70M Operation And Maintenance Manual

Reference pressure monitor with high pressure mounting system
Table of Contents

Advertisement

Quick Links

RPM4/HPMS A70M/A20M-AF™
Reference Pressure Monitor with
High Pressure Mounting System
Operation and Maintenance Manual
NSN 6685-01-529-6990 RN
(2 of 3)
© 2007 DH Instruments, a Fluke Company

Advertisement

Table of Contents
loading

Summary of Contents for DH Instruments HPMS A70M

  • Page 1 RPM4/HPMS A70M/A20M-AF™ Reference Pressure Monitor with High Pressure Mounting System Operation and Maintenance Manual NSN 6685-01-529-6990 RN (2 of 3) © 2007 DH Instruments, a Fluke Company...
  • Page 2 5,142,483; 5,257,640; 5,331,838; 5,445,035. Other U.S. and international patents pending. AutoRange, AutoZ, DH Instruments, DH, DHI, CalTool, COMPASS, GPC1-1000, PGC-1000-AF, RPM4, QDUT, Q-RPT, RPM4 and SDS are trademarks, registered and otherwise, of DH Instruments, a Fluke Company. Document No. 550136b-01...
  • Page 3: Table Of Contents

    O P E R A T I O N ..............1 5 USER INTERFACE ...15 3.1.1 MAIN RUN SCREEN...15 3.1.2 FUNCTION / DATA KEYPAD LAYOUT AND PROTOCOL ...17 3.1.3 REMOTE [ENT] (ENTER) FOOTSWITCH...18 3.1.4 SOUNDS ...18 Page I TABLE OF CONTENTS © 2007 DH Instruments, a Fluke Company...
  • Page 4 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL GENERAL OPERATING PRINCIPLES ...18 3.2.1 PRESSURE READY/NOT READY...18 3.2.2 GAUGE MODE, DYNAMIC COMPENSATION FOR ATMOSPHERIC PRESSURE ...19 3.2.3 MULTIPE RANGES (Q-RPTS, AUTORANGE AND INFINITE RANGING) ...19 3.2.4 AUTOMATED TEST AND CALIBRATION SEQUENCES ...20 3.2.5 HPMS (HIGH PRESSURE MOUNTING SYSTEM) ...20...
  • Page 5: Table Of Contents

    5.4.4.6 DISPLAY ...117 5.4.4.7 COOLING FAN...118 5.4.5 Q-RPT MODULE PNEUMATIC SCHEMATIC...118 RPM4 ILLUSTRATED PARTS BREAKDOWN ...118 HPMS MAINTENANCE... 124 5.6.1 HPMS OVERHAUL ...124 HPMS ILLUSTRATED PARTS BREAKDOWN ...125 TABLE OF CONTENTS Page III © 2007 DH Instruments, a Fluke Company...
  • Page 6 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL T R O U B L E S H O O T I N G ............1 2 9 A P P E N D I X .
  • Page 7: Tables

    Table 18. 8 bit ready status register... 100 Table 19. Program message list ... 101 Table 20. Calibration point sequence for A20M and A70M Q-RPTs ... 110 Table 21. RPM4 Front Panel Illustrated Parts Breakdown (see Figure 15)... 118 Table 22. RPM4 Rear Panel Illustrated Parts Breakdown (see Figure 16) ... 119 Table 23.
  • Page 8 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Figure 1. Typical PGC-10000-AF installation layout... 7 Figure 2. RPM4/HPMS front view ... 8 Figure 3. RPM4/HPMS rear view... 9 Figure 4. RPM4/HPMS side view... 10 Figure 5. RPM4 front panel ... 10 Figure 6. RPM4 rear panel... 11 Figure 7.
  • Page 9: A B O U T T H I S M A N U Al

    RPM4/HPMS A70M/A20M-AF is usually delivered as part of an PGC-10000-AF system which includes a PGC1-10000-AF pneumatic pressure controller and a GB-152-AF gas booster. The PGC1 and the GB-152 have their own Operation and Maintenance Manuals.
  • Page 10 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page VIII...
  • Page 11: I N T R O D U C T I On

    RPM4 setup to cover the range. The HPMS mounts the RPM4 at a convenient viewing angle and contains the hardware to isolate the RPM4’s low pressure Q-RPT (A20M (3 000 psi)) when the high pressure Q-RPT (A70M (10 000 psi) ) is in use.
  • Page 12: Specifications

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL SPECIFICATIONS 1.2.1 GENERAL SPECIFICATIONS Power Requirements: Operating Temperature Range: Storage Temperature Range: Weight: Dimensions: Communication Ports Fuses: Pressure Range: Operating Medium: Pressure Connections: Pressure Limits: PGC-10000-AF SYSTEM Temperature: Relative Humidity: © 2007 DH Instruments, a Fluke Company 85 to 264 VAC, 50/60 Hz, 25 VA max consumption and 12VDC, 1.2 A...
  • Page 13: Pressure Measurement Specifications

    PRESSURE MEASUREMENT SPECIFICATIONS 1.2.2.1 QUARTZ REFERENCE PRESSURE TRANSDUCERS (Q-RPT) RPM4 A70M/A20M is configured with two quartz reference pressure transducers (Q-RPT) modules to measure pressure. The Q-RPTs can measure absolute and gauge pressure. Gauge pressure is defined by offsetting atmospheric pressure and applying dynamic compensation for atmospheric changes using the on-board barometer (see Section 3.2.2).
  • Page 14 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 4...
  • Page 15: I N S T A L L A T I On

    UNPACKING AND INSPECTION RPM4/HPMS A70M/A20M-AF is usually delivered as part of an PGC-10000-AF system which includes a GPC1-10000-AF gas pressure controller and a GB-152-AF gas booster. 2.1.1 REMOVING FROM PACKAGING RPM4/HPMS A70M/A20M-AF is delivered, along with its accessories, in a reusable molded plastic, shipping container with polyurethane inserts to hold it in place.
  • Page 16: Inspecting Contents

    Includes system support software and documentation) PGC-10000-AF SYSTEM RPM4/HPMS A70M/A20M-AF is usually delivered as the reference pressure measurement component of a PGC-10000-AF Pneumatic Gauge Calibrator. The PGC-10000-AF system includes: • RPM4/HPMS A70M/A20M-AF: Reference pressure monitor and high pressure mounting system used as the pressure measuring reference of the calibration system.
  • Page 17: Site Requirements

    Consider the placement of the FOOT SWITCH which may need to be accessed frequently while running calibrations. Support facilities required for RPM4/HPMS A70M/A20M-AF include an electrical power source of 85 to 264 VAC, 47 to 440 Hz.
  • Page 18: Setup

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL SETUP 2.4.1 PREPARING FOR OPERATION To prepare RPM4 for check out and operation: Remove the plastic plug from the RPM4/HPMS rear panel TEST connection. Remove the protective plastic sheet from the front panel display.
  • Page 19: Rpm4/Hpms Rear View

    2. INSTALLATION 2.4.2.2 RPM4/HPMS REAR VIEW Foot switch (remote [ENT]) TEST port (DH500 F) cable connector Figure 3. RPM4/HPMS rear view Page 9 © 2007 DH Instruments, a Fluke Company...
  • Page 20: Rpm4/Hpms Side View

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 2.4.2.3 1. Lo Q-RPT shut off valve 2. RPM4/HPMS A70M/A20M Reference Pressure Monitor 3. Pressure relief valve (3 300 psi) 2.4.3 RPM4 FRONT AND REAR PANELS 2.4.3.1 © 2007 DH Instruments, a Fluke Company RPM4/HPMS SIDE VIEW 4.
  • Page 21: Rpm4 Rear Panel

    11. TEST(+) Pressure Port, Hi Q-RPT 12. ATM Pressure Port, Lo Q-RPT 13. ATM Pressure Port, Hi Q-RPT 14. Product Label (bottom of case) Figure 6. RPM4 rear panel Page 11 © 2007 DH Instruments, a Fluke Company 2. INSTALLATION...
  • Page 22: The Atm Ports Of Rpm4

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL USE THE CORRECT PRESSURE CONNECTORS: The RPM4/HPMS TEST port fitting is a DH500 F (see Section 1.2.1). It is NOT a 1/8 in. NPT F. Never use a fitting other than the corresponding male fitting in these connectors. Damage to the connectors and dangerous failure under pressure could result from using incorrect fittings.
  • Page 23: Power-Up And Verification

    Use [RANGE] to select the Hi Q-RPT and repeat the zeroing process. If the display fails to zero properly for either Q-RPT, RPM4 may need repair. CHECKING ABSOLUTE MODE PRESSURE MEASUREMENT CHECKING GAUGE MODE PRESSURE MEASUREMENT Page 13 2. INSTALLATION If desired, change the © 2007 DH Instruments, a Fluke Company...
  • Page 24: Short Term Storage

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL It is normal for RPM4 to indicate a value other than zero when vented when gauge mode is first entered or ranges are changed, especially if AutoZ is OFF, RPM4 has been OFF for some time or its location has changed.
  • Page 25: O P E R A T I O

    Remote communications are also available via RS-232 or IEEE-488 interfaces (see Section 4). The HPMS mounts the RPM4 at a convenient viewing angle and includes a valve and visual indicators to isolate and protect the RPM4’s 3 000 psi Q-RPT (A20M) when using the 10 000 psi Q-RPT (A70M) (see Section 3.2.5).
  • Page 26: Figure 7. Main Run Screen Display Fields

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL DISPLAY FIELD 1. PRESSURE1 Measured pressure 2. UNIT Unit of measure 3. M Measurement mode 4. h Head pressure indicator 5. z AutoZero indicator 6. RR Active Q-RPT position indicator 7. nn/nn Sequence progress indicator 8.
  • Page 27: Function / Data Keypad Layout And Protocol

    (see Section 3.5). These keys enter numerical values command when editing. Figure 8. Keypad layout ], [ ] and [ ] keys allows up, down, reverse and Page 17 © 2007 DH Instruments, a Fluke Company 3. OPERATION This is...
  • Page 28: Remote [Ent] (Enter) Footswitch

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 3.1.3 REMOTE [ENT] (ENTER) FOOTSWITCH Operating the footswitch is the equivalent of pressing [ENT] on the front panel. The remote ENTER feature can be particularly convenient when running AutoTests (see Section 3.3.10) in which the operator’s hands are not free or attention is on the device under test rather than the RPM4.
  • Page 29: Gauge Mode, Dynamic Compensation For Atmospheric Pressure

    3.2.2 GAUGE MODE, DYNAMIC COMPENSATION FOR ATMOSPHERIC PRESSURE The RPM4 A70M/A20M-AF Q-RPTs are intrinsically absolute but they are also used for gauge measurement mode (see Section 3.3.3, subtracting the value of atmospheric pressure, P using AutoZ (see Section 3.2.2). The AutoZ routine that measures P [AutoZ] whenever RPM4 is in the vented condition.
  • Page 30: Automated Test And Calibration Sequences

    The HPMS (high pressure mounting system) holds the RPM4 at a convenient viewing angle and includes a valve and visual indicators to isolate and protect the RPM4’s 2 000 psi (A20M) Q-RPT when using the 10 000 psi (A70M) Q-RPT. See Figure 10 for an HPMS schematic.
  • Page 31 The HPMS pressure relief valve should only be considered a last resort means of protecting the Lo Q-RPT. Following proper operating procedures, the Lo Q-RPT should never be overpressured. Page 21 © 2007 DH Instruments, a Fluke Company 3. OPERATION...
  • Page 32: Figure 9. Rpm4/Hpms Front View

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 1. RPM4 A70M/A20M-AF Reference Pressure Monitor 2. Lo Q-RPT shut-off valve position LED 3. Lo Q-RPT shut-off valve © 2007 DH Instruments, a Fluke Company 4. Lo Q-RPT active valve position LED 5. Lo Q-RPT active caution LED Figure 9.
  • Page 33: Figure 10. Hpms Pneumatic Schematic

    3. Lo Q-RPT pressure relief valve (3 300 psi) 4. Lo Q-RPT shut off valve Figure 10. HPMS pneumatic schematic Page 23 3. OPERATION 5. RPM4 A70M/A20M-AF Reference Pressure Monitor 6. Connection to RPM4 Hi Q-RPT (A70M) (10 000 psi) © 2007 DH Instruments, a Fluke Company...
  • Page 34: Direct Function Keys Summary

    An RPM4 may have multiple ranges. Each Q-RPT in the RPM4 system has a default range which is its maximum span. The default ranges of the RPM4 A70M/A20M are 10 000 psi (A70M) and 3 000 psi (A20M). Additional ranges, lower than the Q-RPT’s maximum span, may also be created using AutoRange (see Section 3.3.4) or AutoTest (see Section 3.3.10).
  • Page 35 Many RPM4 settings and functions are range specific. See Table 2 for identification of range specific settings. For best metrological performance, vent the RPM4 Q-RPT before changing ranges. Active A20M 3000g/3000a Page 25 © 2007 DH Instruments, a Fluke Company 3. OPERATION DF Lo...
  • Page 36: Unit]

    RPM4 allows the unit of measure in which pressure values are displayed to be changed. RPM4 A70M/A20M-AF supports 13 standard pressure units of measure as well as user defined units. To simplify operation, quick access to six units is made available under the [UNIT] key.
  • Page 37: Mode]

    Measures pressure relative to atmosphere (zero is ambient pressure). Range is from zero gauge to full scale, full scale must be greater than zero. The RPM4 A70M/A20M-AF Q-RPTs measure pressure relative to a sealed vacuum reference and are intrinsically absolute. Gauge mode is accomplished by offsetting of atmospheric pressure with dynamic compensation for atmospheric pressure changes between offsets using an on-board barometer (see Section 3.2.2).
  • Page 38: Table 4. Settings Made By Autorange

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL PRINCIPLE RPM4 A70M/A20M-AF is designed to support the calibration and test of a very wide variety of test ranges in absolute and gauge measurement modes. The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a specific test range and setting measurement parameters appropriately for that specific range.
  • Page 39: Leak Ck]

    The LEAK CHECK function is provided as a means of checking and quantifying the leaks that may be present in the system connected to the RPM4/HPMS TEST port. AutoRange 5000g/5000a Page 29 © 2007 DH Instruments, a Fluke Company 3. OPERATION A70M ] and [ ] cursor...
  • Page 40 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL The principle of the LEAK CHECK function is the measurement of the natural decrease or increase of pressure in a fixed volume over time. The LEAK CHECK function allows a leak check time to be set. The total pressure change and the average rate of change over the leak check time are calculated and displayed.
  • Page 41: Display]

    Blanks out the second line of the display (see Section 3.3.6.7). This function is used when a simple display of pressure measured by the RPM4 active range with minimal additional information is desired. This function is a useful Page 31 © 2007 DH Instruments, a Fluke Company 3. OPERATION...
  • Page 42: Avg (Average)

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL OPERATION To set the DISPLAY function press [DISPLAY] from the main run screen. The display is: The cursor is on the active DISPLAY function. Selecting a DISPLAY function returns to the main run screen with the selected function active.
  • Page 43: Rate

    The AutoTest function (see Section 3.3.10) to AutoRange RPM4 based on DUT characteristics and run an automated test sequence is accessed by pressing [ENT] from the Rate run screen. Page 33 3. OPERATION 3001.51 psi a R 0.01 18Avg Ready indicates that all © 2007 DH Instruments, a Fluke Company...
  • Page 44: Dev (Deviation)

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL OPERATION To activate the Rate DISPLAY press [DISPLAY], <2Rate>. Pressing <2Rate> returns to the main run screen with the Rate DISPLAY active. With the Rate DISPLAY active the main run screen is: Current rate of change of pressure in current pressure unit per second.
  • Page 45: Rpt

    MAIN RUN screen with the inactive Q-RPT displayed on the display’s bottom line. The range of the inactive Q-RPT is always the default range, not an AutoRanged range (see Section 3.2.3). Page 35 3. OPERATION 8998.2 psi a D -1.8 T9000.0 © 2007 DH Instruments, a Fluke Company...
  • Page 46: Hi/Lo

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL With the RPT DISPLAY active the main run screen is: Active Q-RPT display. Inactive Q-RPT display with Ready/Not Ready indicator on far left (<*> indicates Ready). When the RPT display is active, executing a range change to a range on the inactive Q-RPT makes the inactive Q-RPT the active Q-RPT.
  • Page 47: Freeze

    To go to a DISPLAY other than Clean, press [DISPLAY] and make a new DISPLAY choice (see Section 3.3.6). Page 37 3. OPERATION 4435.6 psi a F 4420.7 first 99.76 psi a © 2007 DH Instruments, a Fluke Company Pressing...
  • Page 48: Head]

    [SPECIAL], <3Head> (see Section 3.5.3). As a general rule with RPM4 A20M/A70M-AF, use of the HEAD function to assure in tolerance measurements with RPM4 A70M/A20M-AF is necessary only when the head exceeds 10 in.
  • Page 49: Sds]

    [SPECIAL], <3head> (see Section 3.5.3). 3.3.8 [SDS] PURPOSE The SDS function is not used in RPM4 A70M/A20M-AF. 3.3.9 PURPOSE To run the AutoZ function that rezeros the active Q-RPT between full calibrations. See Section 3.5.1, Principle for a full explanation of the AutoZ function.
  • Page 50: Autoz] In Absolute Mode

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Before running AutoZ in gauge mode, ensure that the pressure applied to the Q-RPT is truly zero gauge (atmospheric pressure). If running AutoZ in gauge mode results in a zero offset that RPM4 considers unusually large, <Confirm 0 gauge P!> is displayed.
  • Page 51 PRINCIPLE) to be std,0 14.03 psi a Pstd,0:0.00 ) in the same unit of measure as the std,0 Old P : 0 Pa offset New P : 2745 Pa offset is entered in the std,0 © 2007 DH Instruments, a Fluke Company...
  • Page 52 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Run AutoZ by COM2 AutoZ by COM2 allows the value of P automatically from a DHI RPMx connected to the RPM4 COM2 communications port. The RPMx is read and the new P To access run AutoZ by COM2 press [AutoZ], <2COM2>. The display is:...
  • Page 53: Ent] (Run Autotest)

    Pressure measurement mode and unit: As specified in the AutoTest initialization. • Display resolution: The DUT tolerance divided by 10 and rounded to the nearest digit, or 1 ppm of Q-RPT span, whichever is greater. Page 43 © 2007 DH Instruments, a Fluke Company 3. OPERATION...
  • Page 54: Autotest Initialization

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL • Stability test for Ready/Not Ready indication(see Section 3.2.1): The DUT tolerance divided by 10, or 2 ppm of Q-RPT span, whichever is greater. • Upper limit (UL): The DUT full scale plus 5 %, or the Q-RPT’s maximum UL, whichever is less (see Section 3.4.4).
  • Page 55: Test Execution

    AutoTest. Page 45 3. OPERATION ENTER to run psi g 0.000 to 5000.0 Pressing [ENT] causes the RPM4 to Set valve for XX RPT Set supply to DUT FS © 2007 DH Instruments, a Fluke Company...
  • Page 56 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL The display is: The top line is the standard RPM4 run screen indicating the current measured pressure (see Section 3.1.1). Exercise the DUT if desired using the top line indication of the measured pressure and press [ENT] when complete.The display is: Standard RPM4 run screen indicating current measured pressure.
  • Page 57: Setup]

    Full scale pressure of the range to be saved in gauge mode and absolute mode if available. This can avoid having to recreate SAVING AN A RANGE ANGE Page 47 3. OPERATION 1range 2res 3stab 4UL 5ATest Save range A70M 4000g/4000a © 2007 DH Instruments, a Fluke Company...
  • Page 58: Deleting Autorange Ranges

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Press [ENT] to save the range or [ESC] to return to the MAIN RUN screen without saving the range. Default (<DF>) Q-RPT ranges cannot be saved. They are permanent. Duplicate AutoRange ranges cannot be saved. A duplicate AutoRange range is a range that has the same measurement mode and full scale and uses the same Q- RPT as a range that has already been saved.
  • Page 59: 3Stab

    The stability setting is range specific. A stability setting made in one range does NOT affect other ranges. Display resltn: 0.0010 %FS Stability limit: 0.5 psi/s Page 49 © 2007 DH Instruments, a Fluke Company 3. OPERATION < and >...
  • Page 60: 4Ul> (Upper Limit)

    The over pressure function executes when a Q-RPT measures a pressure that is greater than 104 % of the Q-RPT’s default span. This is 3120 psi (215.2 MPa) for the A20M Q-RPT and 10400 psi (717.2 MPa) for the A70M Q-RPT. © 2007 DH Instruments, a Fluke Company...
  • Page 61: 5Atest

    The first display is the data file header which can be used to toggle through data files: AutoTest: 1data 2file The data file can be viewed immediately Page 51 © 2007 DH Instruments, a Fluke Company 3. OPERATION When an...
  • Page 62: File

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Full scale of the AutoTest with pressure unit of measure and measurement mode. Number of test points executed/total number of points in the AutoTest. Date and time that the last point in the AutoTest was executed.
  • Page 63 RPM4 settings when the File AutoTest is run (see Section 3.3.10, PRINCIPLE). The display is: Page 53 3. OPERATION Edit File AutoTest # Measurement mode: 1absolute 2gauge 1psi 2MPa 3kPa 4bar DUT full scale: 6000 psi g File1 DUT tolerance: 0.25% FS File1 © 2007 DH Instruments, a Fluke Company...
  • Page 64: Special]

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Enter the tolerance in % of full scale (measurement uncertainty, precision or accuracy as desired) of the DUT to be calibrated using this File AutoTest. The value entered will be used along with the DUT full scale to determine RPM4 AutoRange PRINCIPLE).
  • Page 65: 1Autoz

    OFF in absolute and gauge measurement modes. AutoZ in absolute measurement mode The RPM4 A70M/A20M-AF Q-RPTs are intrinsically absolute. They have an evacuated and sealed reference and naturally measure absolute pressure (difference from vacuum). In absolute measurement mode, the source of P atmospheric pressure, with uncertainty significantly better than that of the Q-RPT that is Z>...
  • Page 66: Table 5. Autoz On And Off

    AutoZ in gauge measurement mode, dynamic compensation for atmospheric pressure The RPM4 A70M/A20M-AF Q-RPTs are intrinsically absolute but they are also used in gauge measurement mode (difference from atmosphere). Gauge measurement mode is achieved by subtracting the value of atmospheric pressure from the Q-RPT’s absolute reading using AutoZ and by dynamically compensating for changes in atmospheric pressure between opportunities for AutoZ to execute (see Section 3.2.2,...
  • Page 67 Pascal (Pa). offset 1off 2view 3edit for the active Q-RPT and measurement mode. offset for the active Q-RPT and measurement mode offset Page 57 © 2007 DH Instruments, a Fluke Company 3. OPERATION abs ON...
  • Page 68: Edit Autoz

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 3.5.1.1 The edit AutoZ function should be used only with great care as entering inappropriate values may result in out of tolerance measurements. operation, the value of the AutoZ offset, P using [AutoZ] (seen Section 3.3.9). Before editing P PRINCIPLE.
  • Page 69: 1Com1, 2Com2

    To run a self test of the RS232 ports (COM1 and COM2), press [SPECIAL], <2remote>, <5RS232test>. Page 59 3. OPERATION © 2007 DH Instruments, a Fluke Company...
  • Page 70: 3Head

    <3User>, is to create a custom liquid with a user entered density. Use [HEAD] to set a head height if desired. 3.5.4 <4SDS> PURPOSE This feature is not used in RPM4 A70M/A20M-AF. 3.5.5 <5PREFS> PURPOSE To access a menu of RPM4 operational preferences and functions.
  • Page 71: 1Scrsvr

    Select between <2lo>, <3mid> or <4hi> to adjust the valid key press tone frequency. Select <1none> to suppress the valid key press tone. 1ScrSvr 2sound 3time 4ID 5level > Page 61 © 2007 DH Instruments, a Fluke Company 3. OPERATION...
  • Page 72: 3Time

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL The sound function only affects the valid key press tone. The invalid key press tone and other RPM4 sounds cannot be adjusted or suppressed. 3.5.5.3 PURPOSE To view and edit the RPM4 internal time and date settings.
  • Page 73: 5Level> (Security)

    It is recommended that the low security level be maintained at all times. If there is a risk of unauthorized changing of the security level, changing authority should be password protected (see OPERATION of this section). Page 63 3. OPERATION Inadvertent, © 2007 DH Instruments, a Fluke Company...
  • Page 74: Table 6. Security Levels

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL The security levels are structured to support typical levels of operation as shown in Table 6. Specifically, the security levels prevent execution of the functions accessed by the key strokes marked by “•”: [RANGE]...
  • Page 75: 6Punit

    2edit password Restrictions: 1none 2low 3medium 4high Select a different level or Password: pppppp 0 disables password RPM4 Password: pppppp factory secondary Page 65 © 2007 DH Instruments, a Fluke Company 3. OPERATION This condition, with a SNnnnn-xx The typical user...
  • Page 76: 7Internal

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL customizes the [UNIT] function key to support his/her most commonly used units. Up to six units can be included in the [UNIT] screen. OPERATION To customize the [UNIT] function key, press [SPECIAL], <6PresU> from the MAIN RUN screen.
  • Page 77: 1Baro

    To view the value of atmospheric pressure as measured by the RPM4 on-board barometer. PRINCIPLE RPM4 A70M/A20M-AF is equipped with an on-board barometer. The atmospheric pressure measurements made by the on-board barometer are used for dynamic compensation of atmospheric pressure when making gauge pressure measurements (see Sections 3.2.2, 3.3.3,...
  • Page 78: 4Lo Vnt

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL When the automated read rate function is ON, three pressure rate of change dependent read rates are used. The result is three display update rates: For situations in which maximum reading precision is desired regardless of pressure rate of change, the RPM4 automated read rate function can be turned OFF.
  • Page 79: 8Cal

    Sets most general operating parameters back to default values. Does not affect calibration coefficients, remote interfaces or AutoRange ranges. The Reset – Sets resets are itemized in Table 9. 1sets 2units 3ATest 4cal 5all Page 69 © 2007 DH Instruments, a Fluke Company 3. OPERATION These...
  • Page 80: Units

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL RESET [UNIT] [MODE] [RANGE] [HEAD] Stability Limit Upper Limit Resolution AutoZ AutoZ Leak Check Screen Saver Key Sounds Lo Vnt ReadRate 3.5.9.2 PURPOSE Sets the pressure units available under the UNIT function to default selections depending (see Section 3.3.2).
  • Page 81: Cal

    Level (Security) Reset security level to low Parallel Measurement Mode Page 71 3. OPERATION RESULT SEE SECTION 5.3.1 5.3.5 5.2.2 5.3.7 3.5.1 3.5.1 RESULT SEE SECTION 3.5.9.1 3.5.9.2 3.5.9.3 3.5.9.4 3.5.2 3.5.2.3 3.5.5.5 3.5.7.3 © 2007 DH Instruments, a Fluke Company...
  • Page 82 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 72...
  • Page 83: Overview

    This pin is the common return for the TxD and RxD signals. IBM PC/XT DB-9F IBM PC/XT DB-9M TO RPM4 DB9F CONNECTIONS DB-25M DB-9F Page 73 4. REMOTE OPERATION The COM1 receive buffer is DESCRIPTION CONNECTION DB-9M DB-9F © 2007 DH Instruments, a Fluke Company...
  • Page 84: Ieee-488

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 4.2.1.2 The RPM4 IEEE-488 interface is located on the back of the unit. The physical and electrical interface conforms to IEEE Std 488.1-1987 Subset E2 and IEEE Std. 488.2-1992. You should not attempt to communicate with the IEEE-488 interface while using the COM1 interface.
  • Page 85: Classic Program Message Format

    Because of this, you can only send one Command program message at a time while using the COM1 port. 4. REMOTE OPERATION Page 75 © 2007 DH Instruments, a Fluke Company However, in RS232...
  • Page 86: Using Query Type Commands

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Examples: • IEEE-488 enhanced mode command series using query operator: “*CLS?” “UNIT? KPA” (Generates a reply. User must wait for reply before continuing) “MMODE? A” (Generates a reply. User must wait for reply before continuing) “PR?”...
  • Page 87: Commands

    Read the RPM4 software version. Read or set the AutoZero Z for the specified Q-RPT. offset Read or set the AutoZ Z for the specified Q-RPT (old command). offset Page 77 4. REMOTE OPERATION “Active” Q-RPT assumed. © 2007 DH Instruments, a Fluke Company...
  • Page 88: Error Messagess

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 4.4.2 ERROR MESSAGESS REPLY ERR# 0 ERR# 2 ERR# 3 ERR# 4 ERR# 5 ERR# 6 ERR# 7 ERR# 8 ERR# 9 ERR# 10 ERR# 11 ERR# 12 ERR# 13 ERR# 14 ERR# 17...
  • Page 89: Program Message Description Overview

    Indicates related commands (“----“) and refers to manual sections giving detail on RPM4 operation corresponding to the program message. 4. REMOTE OPERATION The “Reply” label shows a typical reply to the Page 79 © 2007 DH Instruments, a Fluke Company...
  • Page 90: Program Message Descriptions

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 4.4.4 PROGRAM MESSAGE DESCRIPTIONS Purpose To allow the host PC to communicate with a device connected to the RPM4 COM2 port. Classic “#xx” Arguments Remarks The RPM4 COM2 port can be used to communicate to another RS232 device (such as another RPM4).
  • Page 91 (range of 500 kPa in gauge mode) (Lo Q-RPT is used) (range of 50 psi in absolute mode on the Lo Q-RPT) (No reply if GPIB-488) (Hi Q-RPT’s previous range made active) (No reply if GPIB-488) Page 81 © 2007 DH Instruments, a Fluke Company...
  • Page 92 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL AUTOZEROn Purpose Read or set the status of the AutoZ function. Command “AUTOZEROn state” Query “AUTOZEROn?” Classic “AUTOZEROn= state” “AUTOZEROn” Default “AUTOZERO1” Optional Suffix “n” Arguments ‘state’: Remarks The RPM4 AutoZ function can be turned ON and OFF. There is a separate AutoZ flag for the gauge, absolute, and differential measurement modes for each Q-RPT.
  • Page 93 “9600,N,8,1” Example Sent: “COM2=9600,N,8,1” (classic) Reply: “9600,N,8,1” Errors ERR# 7: Missing or improper program message argument(s). See Also “COM1” 3.5.2.1 4. REMOTE OPERATION (No reply if GPIB-488) (No reply if GPIB-488) Page 83 © 2007 DH Instruments, a Fluke Company...
  • Page 94 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL CONTn Purpose Have the RPM4 reply continuously as each new measurement becomes available. Query “CONTn?” Classic “CONTn” Optional Suffix “n” Remarks This command is somewhat different from most as it enables the RPM4 to continuously output the selected Q-RPT’s measurement without further queries.
  • Page 95 “ID=RPM4 #A01” (classic) Reply: “RPM4 #A01” Errors ERR# 6 The ‘string’ argument was longer than 12 char. See Also 3.5.5.4 4. REMOTE OPERATION (No reply if GPIB-488) (No reply if GPIB-488) Page 85 © 2007 DH Instruments, a Fluke Company...
  • Page 96 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL L2 / L3 Purpose Read or set the type of program command format to use (classic or enhanced). Command “L2” “L3” Defaults “L2” Remarks The user can select the type of remote command format using these simplified commands.
  • Page 97 Errors ERR# 6: Missing or improper program message argument(s). See Also “L2 / L3” 4.3, 3.5.2.3 4. REMOTE OPERATION Absolute mode Gauge mode (No reply if IEEE-488) (No reply if IEEE-488) Page 87 © 2007 DH Instruments, a Fluke Company...
  • Page 98 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL PCALn Purpose Read or set the calibration information for the Hi or Lo Q-RPT. Command “PCALn adder, mult, CalDate” Query “PCALn?” Classic “PCALn= adder, mult, CalDate” “PCALn” Defaults “PCALn = 0.0, 1.0, 19800101” Optional Suffix “n”...
  • Page 99 “R,2306.265 kPa a,0.011 kPa/s” (no barometer) Example Query sent: “QPRR” (classic) Reply: “R,2306.265 kPa a,0.011 kPa/s,97.000 kPa a” See Also “PR”, “PRR”, “SR” 3.1.1, 3.2.1, 3.5.7.1 4. REMOTE OPERATION (no barometer) Page 89 © 2007 DH Instruments, a Fluke Company...
  • Page 100 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL RANGE Purpose Change the active range to the default range of the Hi or Lo. Replies the active range full scale, pressure unit and measurement mode. Command “RANGE Rng” Query “RANGE?” Classic “RANGE= Rng”...
  • Page 101 “0.01” (No reply if GPIB-488) Example Cmd sent: “RES? .01” (enhanced) Query reply: “0.01” Example Sent: “RES=.01” (classic) Reply: “0.01” Errors ERR# 6 The argument is invalid. See Also 3.4.2 4. REMOTE OPERATION Page 91 © 2007 DH Instruments, a Fluke Company...
  • Page 102 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL RESET Purpose Reset the user’s settings to factory defaults. Corresponds to front panel “Reset – Sets”. Command “RESET” Classic “RESET” Remarks The RPM4 has user settings (units, resolution, etc.) that can be reset to factory defaults.
  • Page 103 Reply: “0.10 %” Errors ERR# 6: The argument was invalid. ERR# 10: Invalid or missing command suffix. See Also “SS” 3.4.3 4. REMOTE OPERATION The status is replied when the next Page 93 © 2007 DH Instruments, a Fluke Company...
  • Page 104 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Purpose Read or set the current pressure stability limit. Command “SS limit” Query “SS?” Classic “SS=limit” “SS” Optional Suffix “n” Arguments Limit: Remarks The stability limit can be read and set as a pressure. The stability limit is used as the Ready/Not Ready criteria..
  • Page 105 Sent: “UL=1000” (classic) Reply: “1000.00 kPa a” Errors ERR# 6: The “limit” argument is out of range. See Also 3.4.4 4. REMOTE OPERATION (No reply if GPIB-488) (No reply if GPIB-488) Page 95 © 2007 DH Instruments, a Fluke Company...
  • Page 106 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL UNITn Purpose Read or set the pressure unit of measure unit and measurement mode. Command “UNITn unit (, ref”)” “UNITn unitn (, ref”)” “UNITn unitg (, ref”)” Query “UNITn unita (, ref”)” “UNITn unitd (, ref”)”...
  • Page 107: Status Reporting System

    ) for the specified Q-RPT (Hi or Lo) can be accessed with this offset “ZOFFSET1 2.1, 0, 0” “ 2.10 Pa, 0.00 Pa, 0.00 Pa” “ZOFFSET=97293.1, 3.02, 0” “ 97293.10, 3.02, 0.00” One of the arguments is out of range. Page 97 © 2007 DH Instruments, a Fluke Company 4. REMOTE OPERATION...
  • Page 108: Status Byte Register

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 4.5.2 STATUS BYTE REGISTER The RPM4 contains an 8 bit Status Byte Register that reflects the general status of the RPM4. OPER RQS/MSS (128) (64) This register is affected by the RPM4 reply output queue, the Error Queue, the Standard Event Status register and the Ready Event Status register.
  • Page 109: Standard Event Register

    This bit is not supported as the RPM4 cannot become the active controller in charge. Operation Complete Bit 0 (1) Indicates that the RPM4 has completed all requested functions. 4. REMOTE OPERATION (16) A command error is Page 99 © 2007 DH Instruments, a Fluke Company...
  • Page 110: Ready Status Register

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 4.5.4 READY STATUS REGISTER The RPM4 contains an 8 bit Ready Status Register that reflects specific RPM4 Q-RPT measurement ready events. Enabled events in this register will set or clear the RSB bit of the Status Byte Register.
  • Page 111: Ieee Std. 488.2 Common And Status Program Messages

    PON and QYE bits, the argument would be 128 + 4 = 132. “∗ESE=128”(enables the PON bit) “128” (no reply if IEEE-488) n is not valid. “∗ESR?” “20” (the QYE and EXE bits are set) Page 101 © 2007 DH Instruments, a Fluke Company 4. REMOTE OPERATION...
  • Page 112 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL ∗IDN Purpose Identify the RPM4 version, range, and serial number. Query “∗IDN?” Remarks The identification reply is made up of the manufacture, the model, the serial number and the software version. Each is separated by a comma.
  • Page 113 “RSR?” Remarks The Ready Status Register contents are cleared after reading. The reply is in decimal numeric form. Example Sent: “RSR?” (enhanced) Reply: “6” (The MEAS and NRDY) 4. REMOTE OPERATION Page 103 © 2007 DH Instruments, a Fluke Company...
  • Page 114 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 104...
  • Page 115: Maintenance, Adjustments A N D C A L I B R A T I O

    OVERVIEW RPM4 A70M/A20M-AF was designed for maintenance free operation. No maintenance is required other than: • Regular AutoZeroing of quartz reference pressure transducers (Q-RPTs) (see Section 3.5.1). • Periodic calibration of Q-RPTs (see Section 5.3). • Adjustment of the on-board barometer (see Sections 5.2.2).
  • Page 116: Absolute Mode Autozero

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL If the RPM4 is used only in gauge mode, Autozeroing in absolute mode is not necessary and vice-versa. completely independent one from the other. See Section 3.5.1 for complete information on Autozeroing of RPTs.
  • Page 117: Rpm4 Q-Rpt Calibration

    CalTool for RPTs software supports the calibration process of RPM4 Q-RPTs. CalTool and its documentation are provided on the RPM4 A70M/A20M-AF Support Disk which is delivered with a new RPM4 A70M/A20M-AF. CalTool for RPTs can also be downloaded from www.dhinstruments.com.
  • Page 118: As Received And As Left Data

    AutoZero is used in normal operation. 5.3.2 EQUIPMENT REQUIRED The recommended calibration standard for the RPM4 A70M/A20M-AF Q-RPTs is a DHI PG7202 piston gauge with a DHI RPM4 A100K reference pressure monitor connected to the PG7202 as its external barometer. Contact DHI for additional information.
  • Page 119: Set-Up And Preparation

    Take care to avoid overpressure of the Lo Q-RPT (A20M with maximum pressure of 3 000 psi). The HPMS rear TEST port connects to both the A20M (3 000 psi) and the A70M (10 000 psi) Q-RPTs. Use the Lo RPT Isolation Valve on the HPMS front panel to isolate the Lo RPT when applying pressure to the Hi RPT (see Section 3.2.5).
  • Page 120: Recommended Calibration Point Sequence

    To prevent operation in absolute measurement mode on a Q-RPT that has been calibrated for gauge mode only. PRINCIPLE RPM4 A70M/A20M Q-RPTs are frequently used in gauge measurement mode only (see Section 3.3.3, PRINCIPLE). Q-RPTs that are used in gauge mode only, may be calibrated in gauge mode with a reference that applies gauge pressures.
  • Page 121: Q-Rpt Calibration Using Caltool For Q-Rpts Software

    To calibrate RPM4 Q-RPTs using CalTool software, refer to Sections 5.3.1, 5.3.2, 5.3.3 and 5.3.4 in this manual and then refer to the CalTool for Q-RPTs Software Manual. CalTool for Q-RPTs software and manual are supplied on the RPM4 A70M/A20M Support Disk and can be downloaded from www.dhinstruments.com.
  • Page 122: Q-Rpt Adjustment Without Caltool For Q-Rpts Software

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL To access Q-RPT calibration information viewing or editing press [SPECIAL], <8Cal>. Select the desired Q-RPT. The <1view> selection displays the calibration information fields. The <2edit> function displays the fields and allows them to be edited. The display is: Date of last calibration in YYYYMMDD format.
  • Page 123 = (non-corrected reading • new PM) + new PA Perform additional verification pressure runs as desired. 5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS offset offset Page 113 © 2007 DH Instruments, a Fluke Company was not set to zero,...
  • Page 124: Rpm4 Repair

    RPM4 and/or HPMS repairs. The RPM4 should be calibrated in the HPMS mounting system as the RPM4 orientation may affect its calibration. To remove the RPM4 A20M/A70M-AF from the HPMS High Pressure Mounting System, proceed as follows (numerical references refer to Figure 12): Turn the HPMS upside down.
  • Page 125: Opening And Closing The Rpm4 Enclosure

    8. RPM4 rack mount/HPMS connection 9. TEST port (DH500 F) 10. Hi Q-RPT connection tube 11. Rack mount bracket/rear support bracket connection 12. Rear support bracket 13. Rear support bracket/HPMS connection Page 115 © 2007 DH Instruments, a Fluke Company...
  • Page 126: Subassembly Description And Location

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL web or have difficulty downloading or loading software, contact your DHI representative or a DHI Authorized Service Provider (see Table 23) for assistance. If you believe you have discovered an error or “bug” in RPM4 software, please report it with complete details by email to cal.repair@dhinstruments.com.
  • Page 127: Power Supply Module

    See Section 5.4.5, Figure 14 for a pneumatic schematic of the Q-RPT module configuration. HI Q-RPT MODULE The RPM4 A70M/A20M-AF Hi Q-RPT is an A70M. The A70M has a pressure range of atmosphere to 10 000 psi (70 MPa) absolute, 0 to 10 000 psi (70 MPa) gauge.
  • Page 128: Cooling Fan

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 5.4.4.7 The cooling fan is mounted on the RPM4 rear panel and connected to the main board. It is a 12V brushless DC fan. 5.4.5 Q-RPT MODULE PNEUMATIC SCHEMATIC Figure 14. Pneumatic schematic of RPM4 Q-RPT Module RPM4 ILLUSTRATED PARTS BREAKDOWN Table 21.
  • Page 129: Table 22. Rpm4 Rear Panel Illustrated Parts Breakdown (See Figure 16)

    IEEE-488 assembly Screw lock kit J1 Accessory assembly Adaptor Figure 16. RPM4 Illustrated Parts Breakdown of Rear Panel Page 119 PART NO. 124154 .30C100MXFS D2510M12B JGW25 .25C160MXPS 06AK2D 218001 103285 401420 102073 402195 2CM2-316 © 2007 DH Instruments, a Fluke Company...
  • Page 130: Table 23. Rpm4 Illustrated Parts Breakdown Of Display Assembly (See Figure 17)

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Table 23. RPM4 Illustrated Parts Breakdown of Display Assembly (see Figure 17) # REQUIRED Figure 17. RPM4 Illustrated Parts Breakdown of Display Assembly © 2007 DH Instruments, a Fluke Company DESCRIPTION PART NO. Display assembly...
  • Page 131: Table 24. Rpm4 Illustrated Parts Breakdown Of Internal Assembly (See Figure 18)

    Electronic chassis Screw Figure 18. RPM4 Illustrated Parts Breakdown of Internal Assembly Page 121 5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS PART NO. 123549 401898-06 401966 .30C60KCSS .30NLOCS 122037-Z APCR206CR16WN APO2PV170 3814-7-BK-RL ZCAT1518-0730T 123565 123707 100550-Z © 2007 DH Instruments, a Fluke Company...
  • Page 132: Table 25. Rpm4 Illustrated Parts Breakdown Of Internal Assembly (See Figure 19)

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Table 25. RPM4 Illustrated Parts Breakdown of Internal Assembly (see Figure 19) # REQUIRED Figure 19. RPM4 Illustrated Parts Breakdown of Internal Assembly © 2007 DH Instruments, a Fluke Company DESCRIPTION PART NO. Manifold, A20M...
  • Page 133: Table 26. Rpm4 Illustrated Parts Breakdown Of Electronic Chassis (See Figure 20)

    Cable assembly, DC power Cable assembly, AC power Figure 20. RPM4 Illustrated Parts Breakdown of Electronic Chassis Page 123 5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS PART NO. 401365 SS-5172-200-00 .30C80KCSS .30NLOCS 401745 103284 .30C300SFTS .30NWSFS 401962 401964 © 2007 DH Instruments, a Fluke Company...
  • Page 134: Hpms Maintenance

    Removal of RPM4 from the HPMS is not recommended unless required to perform RPM4 and/or HPMS repairs. See Section 5.4.1 if it is necessary to remove the RPM4 from the HPMS. Any or all of the following items may be included as part of an RPM4/HPMS A70M/A20M- AF overhaul: •...
  • Page 135: Hpms Illustrated Parts Breakdown

    DH200 collar DH500 gland DH500 collar Adaptor Lockwasher, M4 Washer, M4 Nut, M4 Screw, M3 x 10 Screw, M4 x 10 Page 125 © 2007 DH Instruments, a Fluke Company PART NO. 124142 124143 124155 400902-02 122039 122034 124146 124147...
  • Page 136: Figure 21. Hpms Schematic

    RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL Figure 21. HPMS Schematic © 2007 DH Instruments, a Fluke Company Page 126...
  • Page 137: Figure 22. Rpm4/Hpms-Af Assembly

    5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Figure 22. RPM4/HPMS-AF Assembly Page 127 © 2007 DH Instruments, a Fluke Company...
  • Page 138 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 128...
  • Page 139: Table 28. Troubleshooting Guide

    Use PresU function to customize the UNIT function or reset units to default. 3.5.6, 3.5.9.2 Operation is normal. Press any key to resume full screen power. Adjust screen saver time, if desired. 3.5.5.1 Page 129 6. TROUBLESHOOTING SOLUTION © 2007 DH Instruments, a Fluke Company...
  • Page 140 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL SYMPTOM The Q-RPT range does not cover the AutoRange is not selecting the specified AutoRange full scale pressure. desired Q-RPT for the specified range. The specified operating mode is absolute and absolute is OFF for the desired Q-RPT...
  • Page 141 Set active Q-RPT and range to read on desired test port. 2.4.3.2, 3.2.3, 3.3.1 Press [DISPLAY] and set the Display mode to Rate. 3.3.6.2 Page 131 © 2007 DH Instruments, a Fluke Company 6. TROUBLESHOOTING SOLUTION . Rerun AutoZ with a offset...
  • Page 142 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 132...
  • Page 143: Pressure

    7. APPENDIX MULTIPLY BY 1.0 E-02 1.0 E-02 1.0 E-03 1.0 E-05 1.019716 E-01 7.50063 E-03 1.450377 E-04 1.007206 E-06 4.014649 E-03 4.021732 E-03 4.018429 E-03 2.953 E-04 1.019716 E-05 User defined coefficient © 2007 DH Instruments, a Fluke Company...
  • Page 144 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 134...
  • Page 145: Table 30. Dhi Authorized Service Providers

    OVERVIEW Except to the extent limited or otherwise provided herein, DH Instruments, a Fluke Company warrants for one year from purchase, each new product sold by it or one of its authorized distributors, only against defects in workmanship and/or materials under normal service and use. Products which have been changed or altered in any manner from their original design, or which are improperly or defectively installed, serviced or used are not covered by this warranty.
  • Page 146 RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 136...
  • Page 147 The designation of a the Q-RPT with the lower full scale default range in RPM4s with Lo Q-RPT two Q-RPTs. This is the A20M (3 000 psi) in the RPM4 A70M/A20M –AF. Whether pressure is being measured relative to absolute zero or vacuum (absolute mode) Measurement Mode or relative to atmospheric pressure (gauge mode).
  • Page 148 A system to protect Q-RPTs from overpressure made up of isolation and vent valves System) and internal operating logic. SDS is not used in RPM4 A70M/A20M-AF. The difference between FS and the lowest point in a range. For example, the span of a Span 100 kPa FS range in negative gauge mode is nominally 200 kPa (from - 100 kPa to 100 kPa).