Once the analyzer has been installed, it can be configured for your
application. To do this you will:
Set system parameters:
Calibrate the instrument.
Define the three user-selectable analysis ranges, then choose
autoranging or select a fixed range of analysis, as required.
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Before you configure your BDS 3960, these default values are in
effect:
Auto Ranging:
Alarm Relays:
Teledyne Analytical Instruments
LO = 100ppb , MED = 1000 ppb,
Ranges:
HI = 10 ppm, Over-Range = 100
ppm
ON
Defeated, Alarm 1 at 1000 ppb,
Alarm 2 at 100 ppb HI, Not
failsafe, Not latching.
BDS 3960
42