Table 3-4: Range ID Relay Connections

Pin
Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
Range 3 ID Contact
34
Range 4 ID Contact (Over range)
35
Range 4 ID Contact (Over range)
3.4

Electrolyte Refill of BDS Sensor

The BDS sensor is shipped dry. It must be filled with the
electrolyte before operation. The electrolyte is a caustic solution (10%
KOH), supplied in five 50 ml bottles. Review the Material Safety Data
Sheet (MSDS) in Section A-6 before handling the electrolyte.
To refill the BDS sensor:
1. Loosen the thumb screws on the front and slide the drawer
halfway out.
2. Unscrew the sensor cap and disconnect sensor cable from the
BDS sensor.
3. Pour the electrolyte from the five small bottles into a larger
container.
4. Sparge the electrolyte with nitrogen gas at a flow of 100 CCM
for about 1/2 hour then pour into the provided wash bottle.
5. Ref. to Figure 3.5 for the method of adding electrolyte to the
sensor. It is important that as the sensor is being filled with the
electrolyte, filling is accomplished without trapping gas bubbles
in the lower part of the sensor.
6. Squirt electrolyte content into the sensor. Do it slowly until the
bottom parts of the sensor are fully immersed in the electrolyte.
Teledyne Analytical Instruments
Installation
39