Portable Oxygen Monitor
Appendix
A.1 Specifications
Accuracy: +2% of full scale (at constant temperature
Response Time: 90% in less than 8 seconds at 25 °C
Battery Life: Approximately 2000 hr. continuous use in
System Power: 3 AA alkaline batteries.
Sensor Type: Class R17S
Expected Life: 36 months in air
Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33-
Sensor Cable: Retracted: 2 ft / Extended: 10 ft.
Storage Temp. 10-30 °C (continuous), 5-50 °C
Operating Temp: 0-40 °C
Alarm ranges: HI alarm: 19 to 100 Percent
Alarm accuracy: Alarm thresholds are digitally stored and
Teledyne Analytical Instruments
Range: 0-100% Oxygen
and pressure)
a non-alarm condition
mm × 111.5 mm)
(Intermittent)
LO alarm: 18 to 99 %
alarm affectivity is digitally compared
against the value calculated by the
microprocessor. Hence, the accuracy of the
alarms is the same as the accuracy of the unit
as specified above.
Appendix
37