Teledyne 319R Instruction Manual
Teledyne 319R Instruction Manual

Teledyne 319R Instruction Manual

Oxygen detector

Advertisement

Quick Links

M
ODEL
319R
O
D
XYGEN
ETECTOR
I
M
NSTRUCTION
ANUAL
P/N M59608
ECO# 95-416
7/10/95
TELEDYNE BROWN EGINEERING
Analytical Instruments

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 319R and is the answer not in the manual?

Questions and answers

Summary of Contents for Teledyne 319R

  • Page 1 ODEL 319R XYGEN ETECTOR NSTRUCTION ANUAL P/N M59608 ECO# 95-416 7/10/95 TELEDYNE BROWN EGINEERING Analytical Instruments...
  • Page 2 This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
  • Page 3: Table Of Contents

    3.0 Maintenance Replacing the Fuse..................9 Replacing the Sensor ..................10 Z-2 Cell Warranty Conditions ..............11 Appendix Specifications....................13 Spare Parts List ..................... 14 Drawing List ....................15 Material Safety Data Sheet ................16 TELEDYNE BROWN EGINEERING Analytical Instruments...
  • Page 4 TELEDYNE BROWN EGINEERING Analytical Instruments...
  • Page 5: Introduction

    319R Introduction 1.0 Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures. Modular and compact design provide outstanding performance and reliable on-line monitoring and make the 319R virtually maintenance-free.
  • Page 6: Micro-Fuel Cell

    If it is determined that failure is due to faulty material or workmanship the cell will be replaced free of charge. CAUTION: Evidence of tampering or abuse will render the warranty null and void. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 7: Installation & Operation

    The unit is designed to be panel-mounted in a general purpose area, away from moisture and the elements. The unit should be installed at viewing level in a sheltered area. Refer to the Outline diagram (dwg. C-57490) for the physical dimen- sions of the analyzer. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 8: Analysis Unit Connection

    Output Signals Terminals for 4-20 mA, 0-10 VDC, and range identification are labeled on the rear panel of the analyzer, along with high alarm, low alarm, and sensor failure block terminals. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 9: Alarm Terminal Connection

    4. Attach the red lead of the new cell to terminal #1 and the black lead to terminal #2. 5. Replace the cell on the cell holder by pushing and twisting until it strikes bottom. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 10: Setting Alarms

    5. Remember, the System/Cell Failure alarm is factory set at 2% of full scale; should this alarm trigger, the LED to the left of the front panel display will blink. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 11: Calibration

    SPAN button is released. Disabling the alarms allows air to be purged from the sample system without tripping any alarm set below span, or 21%. NOTE: Do not attempt to adjust any alarm setpoints while the alarms are disabled during the 25-second period. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 12 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 13: Maintenance

    4. Remove the bad fuse and replace it with a 5X20 mm 1A, 250V fast acting fuse for AC units (TBE P/N F-1189 ), or 1/2A fast acting fuse for DC units (P/N F-51). Figure 5-1: Replacing the fuse (AC power). TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 14: Replacing The Sensor

    If the cell package is punctured and air permitted to enter, the cell will immediately start to react to the presence of oxygen. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 15: Cell Warranty Conditions

    If it is determined that failure is due to faulty material or workmanship, the cell will be replaced free of charge. CAUTION: Evidence of tampering or abuse will render the warranty null and void. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 16 3.0 Maintenance 319R TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 17: Appendix

    (63.5 mm × 152.4 mm × 63.5 mm) Faceplate: 3.75 "H × 7.0 "W (95.25 mm × 177.8 mm) Probe unit: 4.0 "H × 6.0 "W × 3 "D (101.6 mm × 152.4 mm × 76.2 mm) TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 18: Spare Parts List

    IMPORTANT: Orders for replacement parts should include the part num- ber and the model and serial number of the system for which the parts are intended. Send orders to: TELEDYNE BROWN ENGINEERING Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580...
  • Page 19: Drawing List

    JP2 - connects pins 1 and 2 R52 is 12.5K R30 is 7.5K R35 is 1.5K 25% range modified circuit changes: JP2 - connects pins 1 and 2 R52 is 50K R30 is 28K R35 is 11K TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 20: Material Safety Data Sheet

    Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes. Manufacturer: Teledyne Brown Engineering Analytical Instruments Address: 16830 Chestnut Street, City of Industry, CA 91749 Phone: (818) 961-9221...
  • Page 21 Chronic exposure from the lead contained in this product is extremely unlikely. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 22 If there is liquid around the cell while in the instrument, wear eye and hand protection. Cleanup Procedures: Wipe down the area several times with a wet paper towel. Use a fresh towel each time. Contaminated paper towels are considered hazardous waste. TELEDYNE BROWN ENGINEERING Analytical Instruments...
  • Page 23 NOTE: The above information is believed to be correct and is offered for your information, consideration, and investigation. It should be used as a guide. Teledyne Brown Engineering Analytical Instruments shall not be held liable for any damage resulting from handling or from contact with the above product.