Current Specifications - Milwaukee 2216-20 Repair Service Instructions

Light commercial digital multimeter
Hide thumbs Also See for 2216-20:
Table of Contents

Advertisement

(3) DC Current
Range
6.000A
10.00A
Response Time: < 1.5 sec.
Maximum measurement time: 5 minutes at 10A with at least 20 minutes rest time.
Overload Protection: AC/DC 20A for 10 second.
(4) AC Current
Range
6.000A *
10.00A
* LCD displays 0.000A when measuring < 0.040A.
Frequency Response: 45 ~ 500Hz (Sine Wave)
AC Conversion Type: RMS sensing, RMS indication
Additional Accuracy by Crest Factor (C.F.): Same as ACV.
Max. Crest Factor: Same as ACV.
Response Time of AC Input: < 2 sec. when the LCD reading is bigger than 1000 digits in each
range.
Maximum measurement time: 5 minutes at 10A with at least 20 minutes rest time.
Overload Protection: AC/DC 20A for 10 second.
(5) Resistance
Range
600.0Ω
6.000kΩ
60.00kΩ
600.0kΩ
6.000MΩ
40.00MΩ *
* There is a little rolling less then ±50 digits when measuring > 10.00 MΩ.
Open Circuit Voltage: Approx. 0.6V
Response Time: < 1.5 sec.
Overload Protection: AC/DC 600V for 10 second.
Resolution
0.001A
±(1% reading + 3 digits)
0.01A
Resolution
Accuracy (Sine Wave)
0.001A
±(1.5% reading + 3 digits)
0.01A
Resolution
0.1Ω
0.001kΩ
±(1% reading + 5 digits)
0.01kΩ
0.1kΩ
0.001MΩ
0.01MΩ
±(2% reading + 5 digits)
5
Accuracy
Accuracy