Testing The Voltage Function - Milwaukee 2216-20 Repair Service Instructions

Light commercial digital multimeter
Hide thumbs Also See for 2216-20:
Table of Contents

Advertisement

Testing the Voltage Function

To verify accuracy in the AC and DC voltage ranges, do the following:
1. Turn the rotary switch to "V " position.
2. Connect the Calibrator to the VΩ and COM inputs on the Meter.
3. Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1.
4. Compare the reading on the Meter display with the display reading shown in Table 1.
5. If the display reading falls outside of the range shown in Table 1, the Meter does not meet
specification.
Table 1 AC Voltage Test:
Step
1
2
3
4
5
6
7
8
9
10
11
12
6. Turn the rotary switch to "V " position.
7. Set the calibration for the voltage from step 1 to 4 in Table 2.
8. Compare the reading on the Meter display with the display reading shown in Table 2.
9. If the display reading falls outside of the range shown in Table 2, the meter does not meet
specification.
Table 2 DC Voltage Test:
Step
1
2
3
4
Input
Frequency
500mV
50Hz
500mV
500Hz
500mV
1kHz
5.000V
50Hz
5.000V
500Hz
5.000V
1kHz
50.00V
50Hz
50.00V
500Hz
50.00V
1kHz
500.0V
50Hz
500.0V
500Hz
500.0V
1kHz
Input
Reading
500mV
497.3 to 502.7
5.000V
4.973 to 5.027
50.00V
49.73 to 50.27
500.0V
497.3 to 502.7
Reading
494.7 to 505.3
494.7 to 505.3
489.7 to 510.3
4.947 to 5.053
4.947 to 5.053
4.897 to 5.103
49.47 to 50.53
49.47 to 50.53
48.97 to 51.03
494.7 to 505.3
494.7 to 505.3
489.7 to 510.3
13