Testing The Ampere Function; Testing The Current Function; Testing The Frequency Function - Milwaukee 2216-20 Repair Service Instructions

Light commercial digital multimeter
Hide thumbs Also See for 2216-20:
Table of Contents

Advertisement

Testing the Ampere Function

To verify the accuracy of AC current measurement functions, do the following:
1. Connect the Calibrator to the A and COM inputs on the Meter.
2. Turn the rotary switch to A~.
3. Apply the inputs for steps 1-4 in Table 5.
4. For each input, compare the readings on the Meter display to the reading in Table 5.
5. If the display reading falls outside of the range shown in the Table 5, the meter does not meet
specification.
Table 5 AC Current Test:
Step
1
2
3
4
6. Turn the rotary switch to A .
7. Apply the inputs for steps 1-4 in Table 6.
8. For each input, compare the reading on the Meter display to the reading in Table 6.
9. If the display reading falls outside of the range shown in Table 6, the meter does not meet
specification.
Table 6 DC Current Test:
Step
1
2
3
4

Testing the Frequency Function

To verify the accuracy of the Meter's frequency function, do the following:
1. Connect the Calibrator to the VΩ and COM inputs on the Meter.
Note: The accuracy of the Calibrator's frequency function must be appropriate for the
specified accuracy of the Meter.
2. Set the rotary switch to Hz.
3. Set Function Generator for the square wave voltage and frequency for steps 1-7 of Table 7.
4. Compare the reading on the Meter display with the display reading shown in Table 7.
5. If the display reading falls outside of the range shown in Table 7, the Meter does not meet
specification.
Source
Frequency
5.000A
50Hz
5.000A
500Hz
10.00A
50Hz
10.00A
500Hz
Source
Reading
5.000A
4.922A to 5.078A
5.000A
4.922A to 5.078A
10.00A
9.82A to 10.18A
10.00A
9.82A to 10.18A
Reading
4.922A to 5.078A
4.922A to 5.078A
9.82A to 10.18A
9.82A to 10.18A
15