Table 7 Frequency Test:
Step
Source
1
8.0Vp-p
2
8.0Vp-p
3
8.0Vp-p
4
8.0Vp-p
5
10.0Vp-p
6
230.0Vp-p
7
230.0Vp-p
Level
Waveform
2.00Hz
Square Wave
80.00Hz
Square Wave
800.0Hz
Square Wave
8.000kHz
Square Wave
50.00kHz
Square Wave
10.00Hz
Sine Wave
100.0Hz
Sine Wave
16
Reading
1.98Hz to 2.02Hz
79.90Hz to 80.10Hz
799.0Hz to 801.0Hz
7.990kHz to 8.010kHz
49.93kHz to 50.07kHz
9.97Hz to 10.03Hz
49.93kHz to 50.07kHz