Error Codes

Various error conditions are detected by the tester and are indicated with the
 icon, "Err", and an error number on the primary display. See Table 7. These
error conditions disable testing and, if necessary, stop a running test.
Error Condition
Self-Test Fails
Over-Temp
Fault Voltage
Excessive Noise
Excessive Probe
Resistance
16
Table 7. Error Codes
Code
1
Return the tester to a Fluke Service Center.
2
Wait while the tester cools down.
Check the installation, in particular, the voltage
4
between N and PE.
Switch off all appliances (Loop, RCD
5
measurements) and move the earth stakes
(earth measurement).
Put the stakes deeper into the soil. Tamp
down the soil directly around the stakes. Pour
6
water around the stakes but not at the earth
ground under test.