Specifications - BC Biomedical PLA User Manual

Pacer load adapter

Advertisement

Input Connections:
Output Connections: The output of the Load Adapter connects to a
Load Selection:
Load Impedance:
Enclosure:
IV.

SPECIFICATIONS

Seven banana jacks provide input connections.
Inputs are labeled with the total load impedance
in ohms. The banana jacks mate with standard
size banana plugs (0.166" diameter by 9/16"
long).
paddle input (electrode) of the DA-2000P.
The load provided by the Load Adapter is
selected by connecting the input signal to the
appropriate input jack.
When the Load Adapter is attached to the DA-
2000P, the user is able to select an input load
impedance from the following values: 50Ω,
100Ω, 300Ω, 500Ω, 600Ω, 1000Ω, and 1500Ω.
The 50Ω selection is the load impedance of the
DA-2000P itself. The load resistances have a
tolerance of ±1%, a temperature coefficient of
±100ppm/°C, and low enough inductance to not
interfere
with
performed.
The enclosure is made of a nonconductive
material. It connects directly to one of the DA-
2000P paddle inputs and covers the paddle
input to prevent accidental contact by the user.
4
the
measurements
being