Movement Artifact; Sensor Alignment And Position; Check Out; Figure 8. 97% Spo2 Temperature Dependence - BC Biomedical FINGERSIM PULSE OXIMETER TEST SYSTEM Instruction And Service Manual

Pulse oximeter test system
Table of Contents

Advertisement

B.

Movement Artifact

Analogous to the clinical environment, poorly attached sensors can result in movement between the oximeter
sensor and the FingerSim™ causing erroneous oximeter SpO
sensor to the FingerSim™ so that the emitting and detecting diode are vertically aligned and in direct contact
with the white transmissive surface about ¼" to ½" from the tip of the FingerSim™. Route the sensor cable in
such a way that it is not influencing the sensor attachment to the FingerSim™. For sensors with cables directed
to the back of the finger, route the cable through the notches in the bottom of the holder (see Figure 9).
C.

Sensor Alignment and Position

The light absorption characteristics of the FingerSim™ are specified for light passing directly through the
FingerSim™. Be sure to vertically align the emitter and detector over the white transmissive surfaces about ¼"
to ½" from the tip of the FingerSim™. Misalignment or positioning on the black edges can cause erroneous
readings.
Before each use, visually inspect each FingerSim™ carefully. Do not use the FingerSim™ if the calibration date
marked on the color coded handle has elapsed. Do not use the FingerSim™ if it is cracked and/or leaking fluid.
CAUTION : The SpO
2
appropriate adjustment. Allow at least one hour stabilization at room temperature before using.
Figure 9.
Sensor Cable Routing Through Holder
VI.
simulation by the FingerSim™ is temperature dependent. See Charts 6, 7 and 8 for the
8
and Pulse Rate readings. Be sure to attach the
2

CHECK OUT

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fingersim

Table of Contents