CONTRAINDICATION: Do not use with reflectance or ear clip sensors. Use only with transmittance type,
WARNING:
FingerSims are fragile and must be handled with care, they contain glass.
CAUTION:
Federal law restricts this device to sale by or on the order of a physician.
CAUTION:
The movement of the FingerSim relative to the oximeter sensor may cause erroneous pulse rate
and/or oxygen saturation readings. Use the FingerSim Holder to facilitate pulse generation
without introducing FingerSim™ movement relative to the oximeter sensor.
CAUTION:
Do not use a FingerSim that is cracked or leaking fluid.
CAUTION:
Avoid extended exposure to sunlight.
CAUTION:
The SpO
appropriate adjustment. Allow at least one-hour stabilization at room temperature before using.
CAUTION:
Do not store the FingerSim outside the recommended Long Term Storage Temperature range
(32°F-104°F). NOTE: Temperatures outside this range for short duration are acceptable (for
example during shipping).
CAUTION:
When testing flexible sensors, ensure the emitter and detector are vertically aligned on opposite
sides of the FingerSim.
CAUTION:
Do not use beyond the calibration date.
CAUTION:
No test system can simulate all possible operating conditions a pulse oximeter may encounter. Use
the FingerSim as an adjunct to other indications to determine proper pulse oximeter operation.
CAUTION:
Improper insertion of the FingerSim into the Holder can cause breakage. Insert the Phantom as
shown in Figure 7.
The FingerSim Pulse Oximeter Test System enables the healthcare professional to evaluate pulse oximeter and
sensor function at three simulated light absorption conditions. These absorption conditions are set to simulate a
typical finger at nominally 97%, 90%, and 80% SpO
pulse amplitudes and rates can be simulated.
Before the availability of the BC Biomedical FingerSim, pulse oximetry systems (oximeter plus sensor) were not
easily tested. The oximeter's measurement of the small pulsatile blood component and the interrelationship of the
oximeter calibration curve with the light emitting characteristics of the sensor made a true oximeter system tester
difficult to conceive. The BC Biomedical FingerSim System, when used as an adjunct to other indicators, aids
the healthcare professional in assessing performance of both oximeter and sensor.
CONTRAINDICATION: Do not use with reflectance or ear clip sensors. Use only with transmittance type,
CAUTION: No test system can simulate all possible operating conditions a pulse oximeter may encounter. Use
the FingerSim as an adjunct to other indicators to determine proper pulse oximeter operation.
I.

GENERAL WARNINGS AND CAUTIONS

finger or toe sensors.
simulation by the FingerSim is temperature dependent. See Charts 6, 7 and 8 for the
2
II.
finger or toe sensors.
3

PURPOSE

levels. In addition, a pulse oximeter's response to various
2