Temperature (28 Ii Only); Ac Current - Fluke 27-II User Manual

Digital multimeters
Table of Contents

Advertisement

Temperature (28 II Only)

Range
-200 C to +1090 C
-328 F to +1994 F
[1]
Does not include error of the thermocouple probe.
[2]
Accuracy specification assumes ambient temperature stable to
applies after 2 hour.

AC Current

Function
Range
600.0 A
A ac
6000 A
60.00 mA
mA ac
[4]
400.0 mA
6.000 A
A ac
[5,6]
10.00 A
[1]
AC conversion for the 27 II is ac coupled and calibrated to the rms value of a sine wave input.
[2]
Below 300 counts, add 1 count and the temperature coefficient is 0.15 x (specified accuracy) / C (>40 C).
[3]
AC conversions for the 28 II are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range.
(5 % to 100 % of range) and 10 A range (15 % to 100 % or range).
[4]
400 mA continuous. 600 mA for 18 hr maximum.
[5]
W 10 A continuous up to 35 C. <20 minutes on, 5 minutes off at 35 C to 55 C. >10 A to 20 A for 30 seconds maximum,
5 minutes off.
[6]
>10 A accuracy unspecified.
Resolution
0.1 C
0.1 F
1 C. For ambient temperature changes of
Resolution
Burden Voltage
0.1 A
100 V/ A
1 A
100 V/ A
0.01 mA
1.8 mV/mA
0.1 mA
1.8 mV/mA
0.001 A
0.03 V/A
0.01 A
0.03 V/A
[1,2]
Accuracy
(1.0 % + 10)
(1.0 % + 18)
5 C, rated accuracy
Accuracy
[1,2]
27 II
28 II
(40 Hz – 1 kHz)
(45 Hz – 2 kHz)
(1.5 % + 2)
(1.0 % + 2)
Digital Multimeters
Detailed Specifications
[3]
49