Chapter 3 Features and Functions
General Measurement Configuration
Number of Digits Displayed
Front Panel operation only. See also "Integration Time," on page 58.
You can set the number of digits shown in the display to 4, 5, 6, or 7
full digits, plus a "
You can set both the number of digits displayed and the Integration
Time in a single operation by pressing digits keys (see page 43).
You can change the number of digits displayed without disturbing
the Integration Time settings by pressing the
(see page 43).
The number of digits is set to 6
remote interface reset or preset.
The number of digits shown is dependent upon integration time and
filter settings. The meter will not allow you to show more digits than
the practical measurement capability of the meter. You can, however,
reduce the number of digits shown. The minimum number of digits
1 2
available is 4
. (See the table on page 58).
62
1 2
" digit which can only be a "0" or "1".
1 2
digits at power-on and after either a
<
>
or
keys